Guidance and Manufacturer’s Declaration for Electromagnetic Immunity
and Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and This Device:
The Oxlife Liberty
®
Portable Oxygen Concentrator is intended for use in an
environment where the radiated RF disturbances are controlled. Do not use the
device near active HF surgical equipment and the RF shielded room of an ME
system for magnetic resonance imaging, where the intensity of EM
disturbances is high. Electromagnetic interference can be minimized by
maintaining the distances described below based on the output of the
equipment below.
If the system is compromised due to excessive RF disturbances, an alert or
alarm condition may occur, and the device may power off.
Guidance and
manufacturer’s declaration – electromagnetic emissions
The Oxlife Liberty
®
Portable Oxygen Concentrator is intended for use in the
electromagnetic environment specified below. The customer or the user of the
Oxlife Liberty
®
Portable Oxygen Concentrator should assure that it is used in
such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The Oxlife Liberty
®
Portable Oxygen
Concentrator uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The Oxlife Liberty
®
Portable Oxygen
Concentrator is suitable for use in all
establishments, including domestic, and those
directly connected to the public low-voltage
power supply network that supplies building
used for domestic purposes.
Harmonics
IEC 61000-3-2
Class A
Flicker
IEC 61000-3-3
Complies
RF Emissions
CISPR 14-1
Complies
The Oxlife Liberty
®
Portable Oxygen
Concentrator is not suitable for interconnection
with other equipment.
RF Emissions
CISPR 15
Complies