![O2 Concepts OXLIFE LIBERTY Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/o2-concepts/oxlife-liberty/oxlife-liberty_user-manual_4498906035.webp)
35
Cannula Use
Your physician, healthcare provider or
authorized dealer will recommend the proper
cannula for your use.
Ensure selected cannula is compliant to
ISO 80601-2-69 Medical Electrical Equipment,
particular requirement for basic safety and
essential performance of oxygen concentrator
equipment.
Use of a cannula not specified for use with oxygen may impair the
performance of your device.
DO NOT
use cannula tubing longer than 7 feet (2.13m) in Pulse Flow Mode.
DO NOT
use cannula tubing longer than 50 feet (15.25m) in Continuous Flow
Mode.
A single lumen nasal cannula rated at 6 liters per minute is required to
ensure proper oxygen delivery.
To ensure proper oxygen flow, confirm that the cannula is not pinched or
blocked in any way.
The cannula fitting may be tight.
DO NOT
use grease or oils to lubricate the oxygen outlet port.
Read and follow the instructions included with the cannula and provided by your
authorized dealer.
Regularly clean and replace your cannula as instructed by the manufacturer and
authorized dealer or healthcare provider.
O
2
Concepts
®
recommends the following cannulas by Salter Labs:
For Pulse Mode Use: 7ft (2.13m) cannula: Part #16SOFT-7, or equivalent.
For Continuous Mode Use: 50ft (15.25m) cannula: Part #OTC-203-50,
or equivalent.