41
Invalid Motor
Temperature Alarm
(technical)
(high priority)
When the motor temperature is out of a valid range for
15 seconds, the screen will flash red at 2Hz, sound a
10-beep sequence repeated after 2.5 seconds of silence
and the text
“Invalid Motor Temperature” will flash on the
screen. The alarm will continue to sound until the power
button is pressed momentarily to silence the audio;
or the power button is held to power off the unit;
or 15 seconds has expired at which time the device
will power off.
Solutions
Relocate the device to improve airflow or to
a cooler environment and allow the device to reach
operating temperature. Utilize an alternate source of
oxygen if necessary and contact your authorized dealer.
Invalid Box
Temperature Alarm
(technical)
(high priority)
When the internal box temperature is out of a valid range
for 15 seconds, the screen will flash red at 2Hz, sound a
10-beep sequence repeated after 2.5 seconds of silence
and the text “Invalid Box Temperature” will flash on the
screen. The alarm will continue to sound until the power
button is pressed momentarily to silence the audio;
or the power button is held to power off the unit;
or 15 seconds has expired at which time the device
will power off.
Solutions
Relocate the device to improve airflow or to a cooler
environment and allow the device to reach operating
temperature. Utilize an alternate source of oxygen if
necessary and contact your authorized dealer.
Invalid Flow Alarm
(technical)
(high priority)
When the pulse volume is out of a valid range by 15%
for 60 seconds the screen will flash red at 2Hz, sound a
10-beep sequence repeated after 2.5 seconds of silence
and the text “Invalid Flow” will flash on the screen. The
alarm will continue to sound until the power
button is
pressed momentarily to silence the audio; or the power
button is held to power off the unit; or 15 seconds has
expired at which time the device will power off.
Solutions
Check cannula connection. Repair or replace tubing.
Clean or replace side vents. Move device to ensure
adequate air flow. Ensure device is oriented for
operation, vertically with display visible or
horizontally. Change to another source of oxygen and
contact your authorized dealer.