CAREvent® ALS - User Manual
Tidal Volume
150 - 600 ml
Breaths per Minute
20 – 10
I:E Ratio
1 : 2
Automatic Flow Rate
9.0 - 18.0 L/min
Manual Flow Rate
As per automatic setting
Manual Overide Delay Time
17 - 23 Sec
Demand Breathing Flowrate
0 - 120 L/min
Demand Breathing Triggering Pressure
-2.0 cmH
2
0 / -2.0 mBar
Inspiratory Resistance
-6.0 cmH
2
0 / -5.9 mBar
Expiratory Resistance
6.0 cmH
2
0 / 5.9 mBar
Auto Shut Off Delay Time
5 - 8 Sec
Input Pressure
45 - 87 PSI (3.0 - 6.0 Bar)
Maximum Airway Pressure
60 cmH
2
0 / 58.8 mBar
Operating Temperature
-18
o
C - + 50
o
C / 0
o
F - +122
o
F
Storage Temperature
-40
o
C - +60
o
C / -40
o
F - +140
o
F
Input Connection
9/16” DISS
Patient Connection
15 / 22 mm
Weight
16 OZ (0.45 Kg)(ALS)
14 OZ (0.40 Kg)(BLS)
Size
140 x 63 x 73 mm / 5.5” x 2.5” x 2.9”
Cylinder Duration: CAREvent® ALS
31 ~ 103 minutes*
Cylinder Duration: CAREvent® BLS
69 ~ 138 minutes*
1.4. Performance Specifications
* (Based on an Aluminum “D” size cylinder containing 415 Litres of oxygen.)
(All specifications are subject to a tolerance of +/- 10% except the
I:E Ratio which is subject to a tolerance of +/- 20% and Maximum
Airway Pr0/-10%)