CAREvent® ALS - User Manual
4.2. Cleaning the CAREvent® ALS
Resuscitator and Accessories
Routine cleaning of the equipment should be undertaken to
maintain the equipment in a clean condition.
Reusable patient valve swivel housing and diaphram can be cleaned
using a mild soap solution and disinfected using a legally marketed
commercially available disinfectant, suitable for the application.
Single use patient valves and masks should be discarded after
each patient use and replaced with a new unit.
All other components should be wiped clean with a mild soap
solution. Under no circumstances should the complete unit be
allowed to be soaked or immersed in cleaning solutions.
E
Viewing Window
D
Manual Button
C
Setting Selector
H
Gas Supply
Inlet
G
Resuscitator
Body
B
Diaphragm
A
Patient Valve
Swivel-Housing
F
Pressure Relief
System
Disassembly of the resuscitator (fig 3.)
DESCRIPTION
PROCEDURE
CRITERIA
SCHEDULE
BY
PM
Visual
Inspection
User Manual
Chapter 4.1
Monthly
Checking
Device in work
order, gas
tanks are full, no
missing item
Month
User
PM
Function
Check
Regulator work
pressure limiting
pressure
Within
specification
Every 6
months
User
PM
Level I
Service Manual
Within
specification
Every
year
User
Servicing
Level II
Service
Service Manual
Meet product
specifications
Every
2 years
Manufacturer/
Service Center