B2AV800 Series Surgical Table
IM-B2AV800-E01
5-17
No.
Description
Rotary Knob (#5)
Override Control (#11)
1.
Fuse (neutral)
No.
Description
Up
Down
2.
Fuse (live)
12. Floor Lock
Lock
Unlock
3.
Wire Collect Tie
13. Slide
To head
To leg
4.
Power Cord Socket
14. Height
Up
Down
5.
Manual Rotary Knob
15. Tilt
Left
Right
6.
Foot Control Socket
16. Trendelenburg
Rev. Trend.
Trend.
7.
AC Power Indicator (Green LED)
17. Back
Up
Down
8.
Battery Level Indicator(10 bars: Full, 5
bars: Mid, 2 bars: Low)
9.
(not used)
10. ON/OFF Switch (to actuate movement)
11. Override Control Switch (to control
direction of movement)
The B2AV800 Series tables are equipped with Override System (Figure 5-3) for controlling
table movements in the event of Hand Pendant malfunction. The system consists of an
on/off switch, a manual rotary knob and an override control toggle switch.
1.
Turn knob (# 5, Figure 5-3) to select a desired movement.
2.
Toggle switch (# 11, Figure 5-3) up or down towards the direction of desired movement.
Note:
If the table moves by itself, return the Override Control toggle switch to its
neutral position, which is in the middle, and call for service
Remarque:
Si la table se déplace d'elle-même alors que vous activez le système de
commande de priorité, désactivez-le immédiatement et appelez un
technicien en entretien et en réparation
.
3.
Toggle switch (#10, Figure 5-3) to activate the movement.
4.
Repeat step 1, 2 and 3 as many time as needed to create a desired table position.
Note:
Override System is only suitable for NORMAL patient orientation.
Although table positioning can be performed with the Override System, the problem
should be corrected as soon as possible. Have a qualified service technician repair
the table before further use.
Содержание B2AV800 Series
Страница 2: ......
Страница 18: ...B2AV800 Series Surgical Table IM B2AV800 E01 1 10 This page intentionally left blank...
Страница 32: ...B2AV800 Series Surgical Table IM B2AV800 E01 3 8 This page intentionally left blank...
Страница 42: ...B2AV800 Series Surgical Table IM B2AV800 E01 4 10 This page intentionally left blank...
Страница 50: ...B2AV800 Series Surgical Table IM B2AV800 E01 5 8...
Страница 82: ...B2AV800 Series Surgical Table IM B2AV800 E01 8 2 This page intentionally left blank...
Страница 84: ...B2AV800 Series Surgical Table IM B2AV800 E01 9 2 This page intentionally left blank...
Страница 85: ......