B2AV800 Series Surgical Table
IM-B2AV800-E01
4-6
4.6.1 Performance Checklist
After completing installation procedures (4.1 to 4.5), installation technician shall verify the
performance of table (also refer to 5.2.1
“Hand Pendant Operation”) according to the specifications
and record the results of verification in the chart below:
Test Item
Specification
Result
Verified
Height range of table
without mattress pads
26.8 in ~ 44 in
(68 cm ~ 112 cm)
Slide
12.2 in (31cm)
Kidney Elevator
4.7 in (12 cm)
Trendelenburg
0° ~ 30°
Reverse Trendelenburg
0° ~ 30°
Lateral Tilt
to right 0° ~ 20°
to left 0° ~ 20°
Back Section
up 0°~80°
down 0°~50
o
Leg Section
up 0° ~ 30°
down 0° ~ 90°
left and right 180°
Head Section
up 0° ~ 90°
down 0° ~ 45°
Flex
Back section down 25° max
Seat section down 25° max
Reflex
Back section up 30° max
Seat section up 30° max
Override Switches
Height, Trendelenberg, reverse
Trendelenberg, side tilt, back
section, slide (for
B2AV800LSK, LS, SK, S) floor
locks (for B2AV800 LSK, LS)
Level
+/- 2°
Содержание B2AV800 Series
Страница 2: ......
Страница 18: ...B2AV800 Series Surgical Table IM B2AV800 E01 1 10 This page intentionally left blank...
Страница 32: ...B2AV800 Series Surgical Table IM B2AV800 E01 3 8 This page intentionally left blank...
Страница 42: ...B2AV800 Series Surgical Table IM B2AV800 E01 4 10 This page intentionally left blank...
Страница 50: ...B2AV800 Series Surgical Table IM B2AV800 E01 5 8...
Страница 82: ...B2AV800 Series Surgical Table IM B2AV800 E01 8 2 This page intentionally left blank...
Страница 84: ...B2AV800 Series Surgical Table IM B2AV800 E01 9 2 This page intentionally left blank...
Страница 85: ......