NUVE EN 032 Скачать руководство пользователя страница 5

 

1.

 

INTRODUCTION 

1.1.

 

USE AND FUNCTION 

The EN 032 incubator is designed to provide excellent incubation conditions for applications in 
biology  and  microbiology  laboratories  such  as  medical  and  veterinary  fields;  research  and 
quality  control  examinations  in  pharmaceutical,  food  and  cosmetics  industries  and 
biotechnology. 

The  incubator  provides  temperature  range  between  the  ambient  temperature  plus  5

C  and 

80

C and keeps temperature stable within given tolerances. EN 32 incubators have timer which 

can be programmed up to 99.9 hours including hold position and the incubators stop to operate 
after the set time is completed. Cross contamination risk is minimized by means of natural air 
circulation. 

Stable and homogeneous temperature in the useful volume is provided by means of the forced 
air  circulation  through  the  stainless  steel  heaters  placed  onto  the  three  outer  surface  of  the 
useful volume. 

The insulation proportional to the dimensions of the incubator not only provides a homogenous 
temperature distribution but also makes the operation of incubator economics. The incubator 
ensures  reliable  working  conditions  by  the  programmable  microprocessor  with  high  control 
accuracy. Adjustable safety thermostat offers an additional safety for your operations. 

The EN 032 Incubators are manufactured according to the following standards  
EN 61010-1, EN 61010-2-010, EN 61000-6-3, EN 50419, DIN 12880. 

This device is in compliance with WEEE Regulation. 

If mentioned warnings in this manual are not considered, NUVE will not be responsible from 
their results. 

 

 

 

 

 

 

Содержание EN 032

Страница 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A EN 032 EN 055 EN 120 INCUBATORS INSTRUCTION MANUAL Z14 K25 230 Rev No 03 Rev Date 01 2016...

Страница 2: ...e damaged instrument 4 do not operate the instrument with a damaged cable 5 do not move the instrument during operation In case of a problem contact your N ve agent for an authorized service or mainte...

Страница 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Страница 4: ...UNPACKING 8 5 4 MAINS SUPPLY 8 5 5 POSITIONING 8 5 6 GENERAL PRESENTATION 9 5 7 CONTROL PANEL 10 5 8 PRIOR TO OPERATION 10 6 OPERATING PRINCIPLES 11 6 2 COMPLETION OF THE WORK 12 7 PERIODIC MAINTENANC...

Страница 5: ...natural air circulation Stable and homogeneous temperature in the useful volume is provided by means of the forced air circulation through the stainless steel heaters placed onto the three outer surfa...

Страница 6: ...ard maximum pcs 2 6 2 7 2 10 Power Consumption 300 W 350 W 400 W Power Supply 230 V 50 60 Hz Internal Material Stainless Steel External Material Epoxy polyester painted steel Internal Dimensions W x D...

Страница 7: ...not destroy the structure of the samples Liquids are not heated in sealed containers The boiling points of the samples should be higher than the set temperature Liquids which may expand during heatin...

Страница 8: ...h the instrument please check them 1 ea user s manual and warranty 1 ea electrical connection cable 2 ea shelves 4 ea shelf carriers 5 4 MAINS SUPPLY The instrument requires 230V 50 60 Hz Please make...

Страница 9: ...pace between the equipment and wall 5 6 GENERAL PRESENTATION Figure 1 VENTILATION ADJUSTMENT BUTTON DISPLAY AND CONTROL PANEL DOOR HANDLE PLASTIC PADS ON OFF SWITCH FUSE HOLDER AND SUPPLY INLET VENTIL...

Страница 10: ...Temperature Value Increase and Decrease Keys These keys are used to increase or decrease the values on the temperature display 7 Time Set Key This key is pushed to set the time 1 minute 99 9 hours an...

Страница 11: ...key t n will appear on the temperature display Set the time value by pushing the value increase and decrease keys on the time adjustment side 1 minute to 99 9 hours Hold Push the time set key again t...

Страница 12: ...maintenance 7 2 CLEANING Clean the incubator and at room temperature after disconnecting the power cable Clean the incubator with a wet cloth to remove dirt and dust Use liquid detergent to remove tou...

Страница 13: ...properly The installation of the plug is not defective The mains supply is present If the incubator does not heat check the followings The program is started The safety thermostat value is adjusted hi...

Страница 14: ...14 10 ELECTRICAL CIRCUIT DIAGRAM 10 1 ELECTRICAL CIRCUIT DIAGRAM FOR EN 032 10 2 ELECTRICAL CIRCUIT DIAGRAM FOR EN 055 AND EN 120 HEATERS EN 055 2 x 700 W EN 120 2 x 800 W...

Отзывы: