![NUVE EN 032 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/nuve/en-032/en-032_instruction-manual_1719819011.webp)
11
At most 70% of the surface area of the shelves should be used in order to obtain a uniform
temperature distribution.
6.
OPERATING PRINCIPLES
6.1.
PROGRAMING
Push the temperature set key.
Set the temperature by pushing the value increase and decrease keys on
the temperature adjustment side.
Push the temperature set key again to save the temperature value settings.
Push the time set key.
“t ın” will appear on the temperature display. Set the time value by pushing
the value increase and decrease keys on the time adjustment side.
(1 minute to 99.9 hours, Hold)
Push the time set key again to save the set values.
“dly” will appear on the temperature display. Set the delay time, after
which the program starts, by pushing the value increase and decrease keys
on the time display (1 minute to 99.9 hours or Hold)
Push the time set key again to save the settings.
Push the start / stop key to start the program.
The timer starts to count after the device reaches the set temperature.