22
6.3.
DISINFECTION
Before the disinfection process, please remove all the samples from the
chamber and remove the biological filter in the chamber.
6.3.1.
NüveDis™ DISINFECTION AT 90
⁰
C
Before 90
⁰
C Disinfection is started, fill the humidity tray with distilled water. When
NüveDis™ DISINFECTION selected and approved by pressing the ENTER (F4) key,
warning screens will be seen. Take necessary cautions according to warnings.
There are two types of disinfection
process in device. These are
NüveDis™
90°C disinfection and
optional
nUVeRay™
disinfection
with UV lamp. They can be chosen
by pressing up down arrow and
confirmed by ENTER (F4) button.
Before 90°C Disinfection is started,
security password is shown at the
screen. Authorized persons can
choose by (F1-F2) buttons and
select by (F4).