NUVE EC 160 Скачать руководство пользователя страница 11

11 

 

  Indoor use only 

  A

mbient temperature: 5°C to 40°C 

  Maximum relative humidity for temperature up to 31

°C: 80% 

  Maximum altitude: 2000 m 

  Temperature for m

aximum performance: 15°C / 25°C 

5.2.

  HANDLING AND TRANSPORTATION 

All  handling  and  transportation  must  be  carried  out  by  using  proper  equipment  and 
experienced  staff.  The  instrument  must  be  supported  underneath  and  never  be  turned 
upside down. 

5.3.

  UNPACKING 

Remove  the  cardboard  box  packing  and  the  second  nylon  wrapping  around  the 
instrument.  Ensure  that  no  damage  has  occurred  during  transportation.  The  below 
mentioned are provided with the instrument, please check them; 

 

1 pc. user’s manual and warranty 

  2 pcs. biological filter 

  3 pcs. shelves 

 

2 pcs. of shelves’ support 

 

1 pc. shelves’ stick 

  Humidity tray 

  1 pc. power cable 

  2.5 meters tube for CO

2

 connection 

5.4.

  MAINS SUPPLY 

The CO

2

 Incubator requires 230 V, 50/60 Hz.  

Please make sure that the supplied mains matches the required power ratings which are 
written on the name of plate of the instrument located at the back of the instrument. 

 

Always plug-in the instrument to correctly grounded sockets. 

 

 

A  supply  fitted  with  a  circuit  breaker  should  be  used  for  protection  against 
indirect contact in case of isolation fault. 

5.5.

  POSITIONING 

  Check that the positioning is suitable for the users. 

  Check that the instrument is stable on its four pedestals.  

  Check that the user will be able to follow up the operation even when he deals with 

something else.  

  Check that the positioning of the device prevents interference with other equipment 

in the near surrounding. 

  The bench where the instrument is positioned should be resistant to the weight of the 

instrument and vibration free. 

  Leave at least 30 cm space on the left, right and back side of the instrument. 

Содержание EC 160

Страница 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A EC 160 CO2 INCUBATOR USER S MANUAL Z14 K25 281 Rev No 10 Rev Date 01 2018...

Страница 2: ...ate damaged instrument 4 do not operate instrument with a damaged cable 5 do not move instrument during operation In case of a problem contact your N ve agent for an authorized service or maintenance...

Страница 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...

Страница 4: ...ION 10 5 1 ENVIRONMENTAL CONDITIONS 10 5 2 HANDLING AND TRANSPORTATION 11 5 3 UNPACKING 11 5 4 MAINS SUPPLY 11 5 5 POSITIONING 11 5 6 GENERAL PRESENTATION 12 5 7 CONTROL PANEL 15 5 8 PRIOR TO OPERATIO...

Страница 5: ...RIODIC MAINTENANCE 27 7 2 CLEANING 27 8 DISPOSAL MANAGEMENT CONCEPT 28 9 TROUBLESHOOTING 28 9 1 ERROR CODES 28 9 2 FUSE REPLACEMENT 29 10 OPTIONS 29 10 1 AlerText GSM ALARM MODULE 29 10 1 1 CONNECTING...

Страница 6: ...for reading ambient temperature The volume of EC 160 CO2 Incubator s chamber is 160 liters and is manufactured as in one piece deep drawn stainless steel without any joint in order to prevent contami...

Страница 7: ...ch is harmful for the cells The pH of the culture media is kept constant by sodium bicarbonate The chemical reaction of CO2 and sodium bicarbonate can be represented as CO2 H2O H HCO3 2 H CO3 2 The co...

Страница 8: ...nsitivity 0 1 CO2 Sensor IR Sensor CO2 Sensor Auto Zero Standard In every 12 minutes Safety System Gas expansion thermostat Chamber Capacity 160 liters Number of Shelves std max 3 8 Power Consumption...

Страница 9: ...d be used The set temperature should not destroy the structure of the samples without user s notice The boiling points of the samples should be higher than the set temperature The freezing points of t...

Страница 10: ...WA Belgelendirme Hizmetleri A TOSB stanbul Tuzla Organize Sanayi B lgesi Tepe ren Mevkii 34597 Tuzla stanbul TURKEY Symbol in the operating instructions This symbol refers to important circumstances T...

Страница 11: ...rs tube for CO2 connection 5 4 MAINS SUPPLY The CO2 Incubator requires 230 V 50 60 Hz Please make sure that the supplied mains matches the required power ratings which are written on the name of plate...

Страница 12: ...t be opened frequently If there is laminar flow cabinet in the room it should not blow towards to EC 160 The air ventilation in the room should be minimum If there is air condition in the room it shou...

Страница 13: ...13 Figure 2 9 CO2 MEASUREMENT ACCESS PORT 10 AMBIENT TEMPERATURE SENSOR 11 REMOTE ALARM PORT 12 CO2 TANK PORT 13 BIOLOGICAL FILTER 14 SAFETY THERMOSTAT 15 POWER SWITCH 16 POWER SUPPLY AND FUSES...

Страница 14: ...O2 Measurement Access Port User could use the port in order to connect an external measurement device and measure the CO2 percentage inside the chamber 10 Ambient Temperature Sensor Measure instant am...

Страница 15: ...heir functions correspond to the commands on the display 5 8 PRIOR TO OPERATION 5 8 1 MOUNTING THE SHELVES AND HUMIDITY TRAY Put the shelves supports inside the chamber as it is shown in Figure 4 Figu...

Страница 16: ...s place as in Figure 6 Figure 6 Mounting the biological filter outside of the chamber Mount the biological filter into FILTER port located at the back of the instrument as shown in Figure 2 5 8 3 CO2...

Страница 17: ...se prevent the CO2 gas tank from sun lights and heat The distance between tank and a heat source must be at least 0 5 meters The gas is inside the CO2 tank is heavier than the air and has risks for hu...

Страница 18: ...loaded in order to obtain a uniform temperature distribution 6 2 OPERATING MODE AND MENU 6 2 1 OPERATING MODE NUVE logo appears on the screen for 1 2 minutes for control process This screen is display...

Страница 19: ...stribution Door set temperature equals to chamber set temperature value door temperature difference Lock icon appears on the right bottom corner if the device is locked Press that icon to reach the ke...

Страница 20: ...rature temperature alarm CO2 CO2 alarm and door temperature difference are shown in this screen Pressing up and down button you can increase or decrease value of parameter after choosing with ENTER F4...

Страница 21: ...e and their explanations User orientation aims to make customer usage more interactive and consciously It contains warnings and advices MENU screen are shown by pressing F1 button on operating screen...

Страница 22: ...ON selected and approved by pressing the ENTER F4 key warning screens will be seen Take necessary cautions according to warnings There are two types of disinfection process in device These are N veDis...

Страница 23: ...omes to screen for informing the user Disinfection operation screen Warning page is passed by pressing F4 button and start option is selected by F2 button and confirmed by F4 button Chamber temperatur...

Страница 24: ...t side of the device as in Figure 10 Figure 8 Figure 9 During UV Disinfection process do not to contact with the direct UV light UV lamp should be cleaned frequently against dust and dirt which decrea...

Страница 25: ...ushing ENTER F4 key warning screens will be seen Take necessary cautions according to warnings on the screen Use ENTER F4 key to see next screen UV Disinfection time can be set from 15 minutes to 24 h...

Страница 26: ...al and it contains device settings In this page user can enter user information into user settings Date and time are selected and entered into date and time section Language section is for language se...

Страница 27: ...amage in the chamber Do not use alkaline cleaning reagents and metal brushes for cleaning Use liquid detergent to remove tough dirt Take precautions while handling chemical cleaners Please be aware of...

Страница 28: ...tric electronic parts and disposed of in accordance with applicable local regulations 9 TROUBLESHOOTING If the CO2 Incubator fails to operate please check the followings The power switch is on The fus...

Страница 29: ...fferent mobile phone numbers in case of failures In addition to this message could be sent as an e mail The data is saved on the internal memory of GSM module in case of failures on the communication...

Страница 30: ...NUVE must be informed in advance regarding to SMS message content which will be sent The cost of SIM card and SMS differs according to the GSM network and will be covered by the user 10 2 CO2 CHANGEO...

Страница 31: ...al measurement could be done Also an instrument such as magnetic stirrer can be operated inside This port is factory mounted option The port should be sealed well after the cable is passed through aga...

Страница 32: ...m unit See Figure 12 Mount the remote alarm kit to a wall or board no more than 10 meters away from the device where it is easily accessible 10 6 nUVeRay UV DISINFECTION KIT The disinfection kit is us...

Страница 33: ...33 11 ELECTRICAL CIRCUIT DIAGRAM...

Отзывы: