The menu password for EC 160 CO
2
Incubators is 2112. It is suggested to
change it for your security.
Door temperature value aims for avoiding influence of ambient temperature to
chamber temperature distribution. Door set temperature equals to chamber
set temperature value + door temperature difference.
Lock icon appears on the right
bottom corner if the device is
locked. Press that icon to reach the
keypad for unlocking the device.
Also sensor values button appears
at the left bottom corner.
Security screen contains user
defined password. With the correct
password combination menu screen
opens directly.