9.6 Programming Procedures
Keep the “Pro” button pressed for more than two seconds and then release the button.
“Pr” should display for two seconds and then the display will change to AL 1 and alternate
with the 3 digit set value.
It is possible to program:
o
AL 1
Minimum alarm point expressed in % of oxygen concentration
o
AL 2
Maximum alarm point expressed in % of oxygen concentration
o
FSC
Value expressed in % of oxygen concentration corresponding to the current
output full scale value (20mA). 20mA always correspond to 100 % oxygen
concentration.
o
O
2
c
Oxygen sensor calibration point
At the end of the programming procedure the display will show “End” and the instrument
will display the oxygen content of the gas mix.
Below are the actual screen views of each of the modes.
Program
First Alarm
Second Alarm
21% reading
Full Scale Value
Oxygen Sensor
Calibration