Page 14
www.nuvair.com
7.0 Factory Reset
In case it is necessary to reset the Pro N
2
Alarm Analyzer to the factory settings, power on
the Analyzer by pressing “
On/Off
” and “
Adjust
” buttons simultaneously. The display will flash
“
res
”, then scroll to the programming menu.
The Alarms, FSC value, and O2c value will all need to be checked and/or reprogrammed at
this point.
In case of reset, the instrument will delete all the alarms settings, the full scale value,
and any new conversion value of nitrogen sensor will be reset to factory settings.
Before using again the instrument, it may be necessary to program again the alarm
values, the full scale value, and the nitrogen sensor calibration if changed. All the
analysis concentration shown on the display would be wrong. Wrong nitrogen analysis
may lead to death.
8.0 Maintenance
8.1. Analyzer Care
Analyzers immersed in liquid or stored in wet environments may not operate properly.
This may result in incorrect readings. Incorrect gas analysis may result in personal
injury or death.
Protect the analyzer from excessive shock and impact. Excessive shock and impact
may result in incorrect readings. Incorrect gas analysis may result in personal injury
or death.
Protect the analyzer from exposure to hyperbaric environments. Exposure to
hyperbaric environments may result in incorrect readings. Incorrect gas analysis may
result in personal injury or death.
•
Do not clean Analyzer with anything other than a damp soft cloth.
•
Do not immerse in liquid, leave unprotected outside, or store in a wet environment.
•
Protect Analyzer from excessive shock and impact.
•
Protect Analyzer from excessive exposure to sunlight and extreme temperatures.
•
Do not use the Analyzer in a hyperbaric environment.