Pro H2O Alarm Moisture Analyzer
NUVAIR
Page
10
www.nuvair.com
4.2. Programming Procedures
Keep the “Pro” button pressed for more than two seconds and then release the button.
“Pr” should display for two seconds and then the display will change to AL 1 and alternate
with the 3 digit set value.
It is possible to program:
o
AL 1
First alarm point expressed in ppm of Moisture in concentration
o
AL 2
Second alarm point expressed in ppm of Moisture in concentration
o
FSC
Value expressed in ppm of moisture in concentration corresponding to the
current output full scale value 4-20mA output if available.
o
PrE
ambient pressure expressed in mBar. Default value is 1013
At the end of the programming procedure the display will show “End” and the instrument
will display the moisture in percentage of relative humidity content in the gas mix.
Below are the actual screen views of each of the modes.
Program
First Alarm
Second Alarm
Full Scale Value
Pressure Value of Sensor
End