Pro CO
2
TM
Carbon Dioxide Analyzer
NUVAIR Rev.02.14.12
Page
9
Pro CO
2
Carbon Monoxide Analyzer Manual
The Pro CO
2
TM
can be used to monitor an enclosed space or to analyze a regulated gas sample
flow, the contents of a gas cylinder, or the flow from a regulator:
•
If monitoring an enclosed space, simply remove the Flow Adapter Cap to expose the Sensor
face to the atmosphere and allow 15 seconds for the Display reading to stabilize.
•
If analyzing a gas flow, the Sample Flow Method is the preferred method. The flow rate
must equal 0.5 to 1 L/min at atmospheric pressure (1 bar). To produce this flow, a Flow
Restrictor and Regulator may be required. Contact Nuvair if you need assistance.
4.1. Sample Flow Method of Checking Calibration (Preferred)
Nuvair offers CO
2
gas testing kits see addendum for more information.
Testing Gas must not contain Helium for calibration to succeed.
Step 1. Attach
Flexible Tubing
to Gas Sample
Flow of 0.5 to 1
L/min
Step 2. Verify
that Gas is
Flowing Out
Holes in Flow
Adapter Cap
Step 3. Allow Display
Reading to Stabilize
Step 4. Record
Reading while Gas is
Flowing
Flow Restrictor/
Regulator
Assembly
1000 PPM CO
2
Calibration Gas
Canister
Nuvair
Calibration Gas
Specs:
CO 10 ppm
CO2 1000 ppm
Nitrogen Balance