Pro CO

2

 Carbon Monoxide Analyzer Manual 

4.3. Alarm Setting (AL 1 & AL 2) 

1.) Press the “Prog” button for more than two seconds and then release the button.  

On the display will appear “Pr” for two seconds, then “AL1” will appear and be ready 
for changing the value of the low alarm point. After a second the display will show the 
value of “AL1” CO

2

 ppm current setting. 

2.) The blinking digit shows the cursor position. 
3.) Press the “Prog” button to increase the value (from 0 to 9)  
4.) Press the “Adjust” button to move the cursor to the next digit, the “Adjust” button will 

be used to cycle through the rest of the digits. 

5.) To complete your entry and save the CO

2

 ppm value, press the “On/Off” button. You 

will then automatically jump to “AL 2” programming view.   

6.) Repeat steps 3 through 5 to modify and save the “AL 2” high alarm CO

2

 ppm desired 

value. 

7.) Once programming of “AL 2” is complete you will be in the “FSC” Value Screen and 

ready for programming this value. To jump to the end continue to press the “On/Off” 
button until the end screen appears.  

 

4.4. Full Scale Value Setting (FSC) 

Once the alarms have been set the Pro CO

2

 Analyzer goes to “FSC” view so that you can 

change the analog full scale value. It is not necessary to modify this value which is factory 
set at 1999. However if installing a new sensor the value on the sensor can be entered into 
the “Fsc” setting and must be entered into the “Fct” setting. This is the Carbon Dioxide 
concentration corresponding to 20 mA on the analog output. 4mA is the value at 1999 ppm 
of Carbon Dioxide. This value can be changed in the same manner as the Alarm settings: 
 

1.) Press the “Prog” button for more than two seconds and then release the button.  

On the display will appear “Pr” for two seconds, then “AL 1” will appear. Press the 
On/Off button to cycle through the Alarm settings until you reach the “FSC” view. The 
“FSC” screen and a 3 digit value will alternate for a few seconds and the Pro CO

2

 

Analyzer will be ready for adjusting the “FSC” Value. 

2.) The blinking digit shows the cursor position. 
3.) Press the “Prog” button to increase the value (from 0 to 9) (0-19 on the third value) 
4.) Press the “Adjust” button to move the cursor to the next digit, the “Adjust” button will 

be used to cycle through the rest of the digits. 

5.) To complete your entry and save the FSC value, press the “On/Off” button. You will 

then automatically jump to “End” programming view.    

 
5.0. Threshold Alarms 

The CO2 analyzer will alarm until it has reached the AL1 (low) set point if the AL1 has 
been set. Once the gas is above the AL1 set point the analyzer will not alarm until the 
gas value is over AL2 or under AL1 during the gas analyzation. Should the Carbon 
Dioxide gas trip the low or high alarms (AL1 or AL2) the instrument will go into alarm 
mode and will activate the (optional) relays output (open collector max 100mA) and the 
internal buzzer. The display will show the trespassed alarm and the actual measured 
value. To stop the audible alarm, press any key. In this event the Pro CO

2

 Analyzer will 

remain in alarm mode until the analyzed value reads between the AL1 and AL2 set points. 
The relay output typically is used to shut down the compressor. Nuvair can supply the 
necessary components to adapt your compressor to the relay or provide them at the time 
of install on a new compressor. 
 
 

Содержание Pro CO2 Analyzer

Страница 1: ...Operation Manual Pro CO2 AnalyzerTM Carbon Dioxide Analyzer 02 1 12...

Страница 2: ...mportant safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating this equipment to prevent injury or...

Страница 3: ...peration 8 4 1Sample Flow Method 9 4 2Programming Procedures 10 4 3Alarm Setting 11 4 4 Full Scale Value Setting 11 5 0Threshold Alarms 11 6 0 Powering Off 12 7 0 Factory Reset 12 8 0 Maintenance 12 8...

Страница 4: ...ticular attention to information introduced by the following signal words Indicates an imminently hazardous situation which if not avoided will result in serious personal injury or death Indicates a p...

Страница 5: ...Case includes a Digital Display and controls that are environmentally sealed The Analyzer uses a Flow Adapter Cap and Flexible tubing to deliver sample gas to the Sensor Pressurized gases must be regu...

Страница 6: ...to 24 months depending on usage The Sensor is designed for use at atmospheric pressure 0 P S I The gas mixture to be analyzed must be regulated accordingly and any potential for pressure or vacuum mu...

Страница 7: ...0 P S I Use of gases at higher pressures may result in incorrect readings and may damage the Analyzer Incorrect readings may expose the user to high levels of carbon Dioxide resulting in personal inju...

Страница 8: ...yzer using Calibrated Test Gas Tip You can check the battery life and current temperature by holding Adjust button for 3 seconds The display will alternate from battery life to current temperature Cel...

Страница 9: ...e must equal 0 5 to 1 L min at atmospheric pressure 1 bar To produce this flow a Flow Restrictor and Regulator may be required Contact Nuvair if you need assistance 4 1 Sample Flow Method of Checking...

Страница 10: ...point expressed in ppm of Carbon Dioxide concentration o AL 2 High alarm point expressed in ppm of Carbon Dioxide concentration o FSC Value expressed in ppm of Carbon Dioxide concentration correspond...

Страница 11: ...Carbon Dioxide This value can be changed in the same manner as the Alarm settings 1 Press the Prog button for more than two seconds and then release the button On the display will appear Pr for two se...

Страница 12: ...t it may be necessary to program again the alarm values and the full scale value All the analysis concentration shown on the display could be wrong Wrong Carbon Dioxide analysis may lead to death You...

Страница 13: ...er in a hyperbaric environment 8 2 Battery Replacement Be sure to dispose of spent leaking or damaged Battery properly according to local regulations The following pictures illustrate the steps requir...

Страница 14: ...mediate medical attention if eye contact is made or skin stinging persists 8 4 Handling Sensors Replacement Sensors are supplied in sealed bags Normally Sensors do not present a health hazard Before o...

Страница 15: ...move Screws Step 3 Remove Back Step 4 Disconnect Electrical Connector Step 5 Remove Old Sensor from Case by Unscrewing Step 6 Replace with New Sensor Step 7 Remove Shorting Plug from Sensor Electrical...

Страница 16: ...semblies to regulate the gas 10 0 Troubleshooting SYMPTOM REASON SOLUTION Battery symbol Low Battery Change the battery No display Switched off Bad connection Low Battery Switch on Check display batte...

Страница 17: ...nections The relay output uses a special Mini Jack plug that can be purchased separately from Nuvair Insert the plug into the output jack Connections diagram below for additional wiring The plug shoul...

Страница 18: ...Years Power 9 Volt Battery or optional 110 230v electric Response Time Less Than 50 Seconds to 90 of Final Value Stabilization Time 10 Minutes Max Accuracy7 Operating Temperature 41 to 104 F 5 to 40...

Страница 19: ...purchase the warranty will begin with the date of manufacture by NUVAIR Maintenance Items Any materials which are consumed or otherwise rendered not warrantable due to processes applied to them are co...

Страница 20: ...Nuvair 2012 All Rights Reserved Phone 805 815 4044 Fax 805 815 4196 2949 West 5th Street Oxnard CA 93030 USA Email info nuvair com Web www nuvair com Revision 0212...

Отзывы: