GB
- 49 -
Accuracy:
Ear mode:
±0.2 °C/0.4 °F in the measuring range 35.5~42.0 °C
(95.9~107.6 °F)
±0.3 °C / 0.5 °F outside this range
Clinical Bias: -0.5 ~ -0.7 °C
Limits of Agreement: 1.02
Repeatability: 0.19 °C
Temple mode:
±0.2 °C/0.4 °F in the measuring range 35.5~42.0 °C
(95.9~107.6 °F)
±0.3 °C / 0.5 °F outside this range
Clinical Bias: -1.3 ~ -1.5 °C
Limits of Agreement: 1.47
Repeatability: 0.22 °C
Surface temperature mode: ±0.3 °C / 0.5 °F in range 22.0~42.2 °C (71.6~108.0 °F),
otherwise ±4%(however at least ±2.0 °C / 4.0 °F)
Atmospheric pressure:
800~1013 hPa
Protection class:
BF
Degree of protection:
IP22