EMC Table

Guidance and manufacture’s declaration – electromagnetic emission

Emission test

RF emission

Harmonic

Voltage

Group 1

Class B

Not applicable

Not applicable

The Digital Blood Pressure Monitor use RF energy only for 

its internal function. Therefore, its RF emissions are very low 

and are not likely to cause any interference in nearby 

electronic equipment.

The Digital Blood Pressure Monitor is suitable for use in all 

establishments other than domestic and those directly 

connected to the public low-voltage power supply network 

that supplies buildings used for domestic purposes.

CISPR 11

RF emissions

emissions

fluctuations/

flicker emissions

IEC 61000-3-3

IEC 61000-3-2

CISPR 11

Compliance

Electromagnetic environment – guidance

The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment 

specified below. The customer of the user of the Digital Blood Pressure Monitor should 

assure that it is used in such an environment.

Guidance and manufacture’s declaration

Guidance and manufacture’s declaration – electromagnetic immunity

Immunity test

Power frequency 

magnetic field IEC 

61000-4-8

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

3V/m

3V/m

Floors should be wood, concrete or 

ceramic tile. If floor are covered with 

synthetic material, the relative humidity 

should be at least 30%.
Power frequency magnetic fields should 

be at levels characteristic of a typical 

location in a typical commercial or 

hospital environment.

(50Hz/60Hz)

Electrostatic

IEC 61000-4-2

discharge (ESD)

IEC 60601 test  Compliance

level

level

Electromagnetic environment - guidance

The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment 

specified below. The customer or the user of Digital Blood Pressure Monitor should assure 

that it is used in such an environment.

NOTE   UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacture’s declaration – electromagnetic immunity

The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment 

specified below. The customer or the user of Digital Blood Pressure Monitor should assure 

that it is used in such an environment.

27

Compliance

level

EMC Table

Immunity test

80 MHz to 800 MHz
800 MHz to 2.5 GHz

Portable and mobile RF communications 

equipment should be used no closer to any 

part of the Digital Blood Pressure Monitor, 

including cables, than the recommended 

separation distance calculated from the 

equation applicable to the frequency of the 

transmitter.

Where P is the maximum output power 

rating of the transmitter in watts (W) 

according to the transmitter manufacturer 

and d is the recommended separation 

distance in metres (m).

Field strengths from fixed RF transmitters,

as determined by an electromagnetic site 

IEC 60601 test 

level

Electromagnetic environment - guidance

Recommended separation distance

80 MHz to 2.5

GHz

3V/m

3V/m

survey, should be less than the compliance 

level in each frequency range.

b

Interference may occur in the vicinity of 

equipment marked with the following 

symbol:

NOTE 1       At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2       These guidelines may not apply in all situations. Electromagnetic propagation is 

affected by absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 

telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 

broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 

environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the Digital Blood Pressure 

Monitor is used exceeds the applicable RF compliance level above, the Digital Blood 

Pressure Monitor should be observed to verify normal operation. If abnormal performance is 

observed, additional measures may be necessary, such as reorienting or relocating the 

Digital Blood Pressure Monitor.

Radiated RF
IEC 61000-4-3

a

b

a

Over the frequency  range 150 KHz to 80 MHz, field strengths should be less than 3V/m.

Содержание BA2115

Страница 1: ...Model No BA2115 A01 BA2115...

Страница 2: ...ols Use 2 Important Safety Notes 3 K now Your Unit 4 5 IHB Introduction 6 Before Taking A Measurement 7 Quick Reference Guide 8 Battery lnstallation Replacement 9 Adjusting Settings 1 To set year 10 2...

Страница 3: ...sing measured results may be dangerous Follow the instructions of your healthcare provider Operate the unit only as intended Don t use for any other purpose The unit is intended to use in measuring bl...

Страница 4: ...to Air Socket Metal Arm Cuff Warning 1 Do not use rechargeable batteries 2 When using optional AC adapter the AC adapter must comply with the requirements of standard IEC60601 1 3 Optional AC adapter...

Страница 5: ...d bathing for 30 minutes before taking a measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 The cuff can be applied to your left or right wrist Always measur...

Страница 6: ...ve polarities match the polarities of the battery compartment as indicated Remove the battery cover from the battery compartment Replace the battery cover Press the START STOP button All symbols appea...

Страница 7: ...inue with next step Then set the day the time system conversion the hour and the minute by using the same way Continue with step 3 2 To set date and time 10 Press the MEM botton to increase one year P...

Страница 8: ...START STOP button to save the current setting and turn off 5 To set user mmHg display kPa display Adjusting Settings 12 Adjusting Settings 13 Press the SET button and the mark of a user will appear wh...

Страница 9: ...ug is inserted properly in the main unit 1 15 Proper Use of the Arm Cuff 8 Relax your arm keep the palm upward and the fingers curving naturally then turn on the unit and start to measure Wrap the cuf...

Страница 10: ...rts to inflate automatically Press the START STOP button 1 In order to accelerate the measurement time for the systolic pressure more than 180 mmHg press and hold the memory button until the unit infl...

Страница 11: ...before reading the memory To read the average value Press the MEM button The average value symbol will display the average value of the latest 3 times measurement taken Turn to next The 1st latest mea...

Страница 12: ...ower value represents the blood pressure produced by relaxation of the heart muscle Blood Circulation 21 About Blood Pressure Note When a person s systolic and diastolic pressures fall into different...

Страница 13: ...arm and body and keep quiet Error Indicators 22 The following symbol will appear on the display when measuring abnormally 23 Care and Maintenance Do not clean the body and cuff with naphtha thinner or...

Страница 14: ...overvoltage II pollution degree 2 The risk of patient and user can be lowered to acceptable level The unit might not meet its performance specification if stored or used outside the following specifie...

Страница 15: ...lood Pressure Monitor is intended for use in the electromagnetic environment specified below The customer or the user of Digital Blood Pressure Monitor should assure that it is used in such an environ...

Страница 16: ...e does not cover damage incurred by use of the unit not in accordance with the instructions accidental damage or being tampered with or serviced by unauthorized service agents 29 28 EMC Table Recommen...

Отзывы: