Immunity test

80 MHz to 2.5

GHz

3A/m

3A/m

survey,a should be less than the 
compliance level in each frequency range.b
Interference may occur in the vicinity of 
equipment marked with the following 
symbol:

NOTE 1       At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2       These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV 
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the Digital Blood Pressure 
Monitor is used exceeds the applicable RF compliance level above, the Digital Blood 
Pressure Monitor should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the 
Digital Blood Pressure Monitor.

Radiated RF

IEC 61000-4-3

IEC 60601 test  Compliance 

level

level

Electromagnetic environment - guidance

                                Recommended separation distances between 
portable and mobile RF communications equipment and the Digital Blood Pressure Monitor. 

output power 
of transmitter

For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in metres (m) can be estimated using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmitter 
in watts (W) according to the transmitter manufacturer.

Separation distance 

 (m)

according to frequency of transmitter

Pressure Monitor is intended for use in an electromagnetic environment in which radiated 
RF disturbances are controlled. The customer or the user of the Digital Blood Pressure 
Monitor can help prevent electromagnetic interference by maintaining a minimum distance 
between portable and mobile RF communications equipment (transmitters) and the Digital 
Blood Pressure Monitor as recommended below, according to the maximum output power 
of the communications equipment.

80 MHz to 800 MHz

0.117

0.233

0.369

0.738

1.167

2.333

3.689

7.379

11.667

23.333

80 MHz to 800 MHz

0.01

0.1

1

10

100

Содержание BA2015

Страница 1: ...Model No BA2219 A00...

Страница 2: ...3 Konw Your Unit 4 5 IHB Introduction 6 Before Taking A Measurement 7 Quick Reference Guide 8 Battery lnstallation Replacement 9 Adjusting Settings 1 To set year 10 2 To set date and time 10 11 3 To...

Страница 3: ...or moderate injury to the user or patient or damage to the equipment or other property Read all information in the instruction manual and any other literature included in the box before using the uni...

Страница 4: ...Hose Air Plug Insert to Air Socket Metal Arm Cuff Warning 1 Do not use rechargeable batteries 2 When using optional AC adapter the AC adapter must comply with the requirements of standard IEC60601 1 3...

Страница 5: ...d bathing for 30 minutes before taking a measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 The cuff can be applied to your left or right wrist Always measur...

Страница 6: ...ative polarities match the polarities of the battery compartment as indicated Remove the battery cover from the battery compartment Replace the battery cover Press the START STOP button All symbols ap...

Страница 7: ...the month and press the SET button and continue with next step Then set the day the time system conversion the hour and the minute by using the same way Continue with step 3 2 To set date and time 10...

Страница 8: ...the START STOP button to confirm user and turn off You can stop the setting anytime when you press the START STOP button to save the current setting and turn off If select to display the date and time...

Страница 9: ...plug is inserted properly in the main unit 1 15 Proper Use of the Arm Cuff 8 Relax your arm keep the palm upward and the fingers curving naturally then turn on the unit and start to measure Wrap the...

Страница 10: ...uff starts to inflate automatically Press the START STOP button 1 In order to accelerate the measurement time for the systolic pressure more than 180 mmHg press and hold the memory button until the un...

Страница 11: ...memory To read the average value Press the button The average value symbol will display the average value of the latest 3 times measurement taken Turn to next The 1st latest measurement value The 2nd...

Страница 12: ...nts the blood pressure produced by relaxation of the heart muscle Blood Circulation 21 About Blood Pressure Note When a person s systolic and diastolic pressures fall into different categories the hig...

Страница 13: ...ve your arm and body and keep quiet Error Indicators 22 The following symbol will appear on the display when measuring abnormally 23 Care and Maintenance Do not clean the body and cuff with naphtha th...

Страница 14: ...y using the oscillometric method The unit is intended for usingin only adult population not applied to the other populations such as neonatal baby It can t be used while the wirst arm has bleeding or...

Страница 15: ...60601 1 2 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnetic compatibility Requirements and tests EN1060 1 Non in...

Страница 16: ...1000 4 8 6 kV contact 8 kV air 6 kV contact 8 kV air 3A m 3A m Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at least 30 P...

Страница 17: ...be necessary such as reorienting or relocating the Digital Blood Pressure Monitor Radiated RF IEC 61000 4 3 IEC 60601 test Compliance level level Electromagnetic environment guidance Recommended sepa...

Отзывы: