Following a manual release, the wrench must be removed from its housing to prevent accidents related to the rotation of the
wrench during the activation of the locking mechanism. In addition, the security indicator label must be replaced during a safety
check carried out by an authorized technician.
6.2.9 Precooling
On the Refrigerated model, a "Pre-cooling" cycle prior to centrifugation allows to cool the centrifuge down to
4 degrees Celsius so it is ready for temperature sensitive samples.
Visual
Instruction
Control
Access to the list of programs by pressing
this icon :
Select the pre-cooling program:
Rotor: Standard Precool
This program is set to cool the centrifuge
and accessories down to 4°C during a 22
minute cycle at 2500 RPM. Acceleration and
Deceleration slope are both set on 9 and
cannot be modified.
Air: Smart Precool
This program is set to cool the centrifuge
until the probe temperature reaches 4⁰C and
will stop when the temperature is reached.
*Select Rotor or Air in user preferences
Start the cycle by pressing
The temperature inside the bowl will show
on the display