10.1.4 Maintenance
The output of an ultraviolet light deteriorates with burning age. The useful life of the light is approximately 5000
hours under specific test conditions.
Note:
Before testing with lamp off, the light may be cleaned with a lint-free cloth dampened with alcohol
or ammonia and water.
The control system monitors UV light usage hour and when the 5000 hour point is reached. The UV light LED
indicator will blink fast along with the display message “Replace UV light”. To view hours or reset timer, perform
the following:
•
Press and hold ENT key for 3 seconds until the menu appears.
•
Press ↑ or ↓ keys to
select UV lamp hours.
•
Press ENT key to highlight hours.
Press ↑ or ↓ key simultaneous until hours zero.
•
Press ENT key to accept cleared hours.
•
Press
key to return to run display.
Main Menu
Time Menu
UV Lamp Hours 3356
Temperature 75.0
Blower On/Off Password none
Содержание LabGard D NU-543-400
Страница 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Страница 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Страница 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Страница 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Страница 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Страница 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Страница 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Страница 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Страница 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Страница 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Страница 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Страница 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...
Страница 69: ...OM0283 Page 69 of 70 Rev 2 July 2021 ...
Страница 70: ...OM0283 Page 70 of 70 Rev 2 July 2021 ...