OM0283
Page 51 of 70
Rev 2 July/2021
7.9 Cleanliness Classification Test for Pharmacy Application
If this cabinet is going to be used within pharmacy, per USP797
1
, the cabinet must be tested to assure compliance to ISO
14644-1:1999, Cleanrooms and Associated Controlled Environments, Part 1: Classification of Air Cleanliness
2
. The
cleanliness classification test is performed using a particle counter to measure particle counts within the cabinet workzone.
Turn on cabinet and let warm up for several minutes. Turn on particle counter and flush out sample tubing line to remove
latent particles. Set the particle counter to measure 0.5 micron or larger particles at the appropriate measuring rate.
“Operational Particle Count Test
3
”
Position the particle counter isokinetic probe at a point 6 inches (152mm) upstream of the aseptic manipulation area (hand
convergence point) and mounted so as not to interfere with the operator’s hand movement. The pharmacy operator will
simulate IV manipulation during the particle count test using non-hazardous materials. A minimum of three (3) 1-minute
particle counts shall be sampled and recorded while the user simulates aseptic compounding manipulations.
“At Rest Particle Count Test”
Take 5 test points in 1-minute intervals on a grid, in a horizontal plane as measured approximately 6 inches (152mm) above
the worksurface. The grid location is designed as the workzone center point and each corner measured 6 inches (152mm)
from the inside perimeter.
Record the 5 particle count values for each of the test points over the 1-minute sample time.
All final count particle concentrations and calculated 95% upper confidence limit shall not exceed 3520 particles per cubic
meter (ppcm) or 100 particles per cubic feet (ppcf).
1
USP28-NF23: United Stated Pharmacopeial Convention, Inc., 12601 Twinbrook Parkway, Rockville, MD
20852, USA, www.usp.org.
2
ISO 14644-1:1999 Cleanrooms and Associated Controlled Environments-Classification of Air Cleanliness, International Organization for Standardization,
Case Postale 56, CH-1211 Geneve 20, Switzerland
3
CAG-002-2006: CETA Compounding Isolator Testing Guide, Controlled Environment Testing Association, 1500 Sunday Drive, Suite 102, Raleigh, NC
27607, USA, www.cetainternational.org
Содержание LabGard D NU-543-400
Страница 7: ...OM0283 Page 7 of 70 Rev 2 July 2021 ...
Страница 9: ...OM0283 Page 9 of 70 Rev 2 July 2021 ...
Страница 10: ...OM0283 Page 10 of 70 Rev 2 July 2021 ...
Страница 11: ...OM0283 Page 11 of 70 Rev 2 July 2021 ...
Страница 12: ...OM0283 Page 12 of 70 Rev 2 July 2021 ...
Страница 16: ...OM0283 Page 16 of 70 Rev 2 July 2021 ...
Страница 21: ...OM0283 Page 21 of 70 Rev 2 July 2021 ...
Страница 34: ...OM0283 Page 34 of 70 Rev 2 July 2021 ...
Страница 65: ...OM0283 Page 65 of 70 Rev 2 July 2021 ...
Страница 66: ...OM0283 Page 66 of 70 Rev 2 July 2021 ...
Страница 67: ...OM0283 Page 67 of 70 Rev 2 July 2021 ...
Страница 68: ...OM0283 Page 68 of 70 Rev 2 July 2021 ...
Страница 69: ...OM0283 Page 69 of 70 Rev 2 July 2021 ...
Страница 70: ...OM0283 Page 70 of 70 Rev 2 July 2021 ...