Page 8 of 12
General care, maintenance and repair
Have your PODIAVAC checked if the corresponding Service LED (9) lights up.
Cleaning the device
Attention!
Always pull the mains plug before cleaning the device. Risk of electric shock!
Clean the housing (incl. the hand-piece and the hose) with a damp (but not wet) cloth.
Do not use any cleaning agents which contain alcohol or solvents or which are aggressive.
Exchange the dust bag at least once a week (more frequently in case of higher load). Do not fill
the dust bag to more than 70 per cent of its capacity. Otherwise, dust may leak out. This, in turn,
reduces the service life of your motor.
Attention: Never use the device without a dust bag!
The suction chamber has to be kept clean at all times.
Replacing the dust bag
Pull the hand-piece connector from the suction chamber (1) towards the front and, afterwards,
pull the dust bag from the pipe socket.
Push the new dust bag (2) onto the pipe socket until it reaches the stop point.
Put the connection of the hand-piece with the dust bag into the suction chamber without jamming
the dust bag in the opening of the suction chamber or damaging it.
The guide pins of the connection of the hand-piece have to be inserted into the guiding holes of
the suction chamber. Make sure that the contact pin is not damaged by any jamming which may
occur.
Cleaning the hand-piece
NEVER clean the hand-piece using boiling water, chemicals, wire brushes or ultra-sound.
NEVER immerse the hand-piece in a disinfectant or ultrasonic bath.
NEVER place or immerse the hand-piece in a liquid, otherwise this might damage the hand-piece.
Make sure that no liquid gets into the motor of the hand-piece.
Twist suction sleeve against hose mounting (unlock bayonet lock) and then pull it off the motor of the
hand-piece in an axial direction (towards the front).
Remove contamination on the suction sleeve, the motor housing and the collet chuck with a brush or
a soft cloth.
The suction sleeve can also be cleaned in an ultrasonic bath.
Once cleaning has taken place, push the suction sleeve on the motor casing again. In this process
make sure that the motor is inserted into the suction sleeve in a centered position.
(See Ill. 2)
Caution!
The snap-fit hooks on the hose mounting have to be inserted into the grooves of the
moulded rubber part. Finally, the suction sleeve and the hose mounting have to be twisted again until