Installation

 

 

Connection the hand-piece: 

 

Push a dust bag onto the pipe socket of the handpiece-connector until it reaches the stop.  

 

Afterwards,  insert  the

 

handpiece-connector  with  the  dust  bag  placed  on  it  into  the  suction 

chamber (see Illustration above - 1).  

In  this  process,  the  contact  pins  on  the

 

handpiece-connector  have  to  be  positioned  pointing 

up and fit into the contact plug on the suction chamber. This is ensured by two guide pins on 
the

 

handpiece-connector. Make sure that the handpiece-connector is in a straight position and 

that it does not become jammed while you insert it. 

Inserting the tool: 

 

Insert  the  tool,  e.g.  a  suitable  grinder,  into  the  tool  holding  fixture  until  it  reaches  the  stop 
point.  Make  sure  that  the  tool  is  seated  firmly  in  order  to  prevent  any  loosening  of  the  tool 
during work.  

 

Connection the foot switch (optional): 

 

Insert the plug of the foot switch into the connector on the right side of the device below the 
fixing hole for the hand-piece. 

Mains supply: 

 

Connect the supplied power cable with the power inlet on the right side of the device. 

Attention!

    Check  the  voltage  (V)  specified  on  the  label  before  connecting  the  device  to  the 

supply voltage. 

 

Plug the power cord into a wall outlet with adequate protection contact. 

The device is now ready for operation. 

 

Warranty 

 
The term of the warranty for Podiavac amounts to 1 year or a limit of 500 hours of operation, except 
wearing parts such as bearings, carbon brushes, dust bag and seals. 
The warranty extended for a period of another 500 operating hours up to a maximum of 1 year, 
if the recommended maintenance of the device was done in time. 
 
 
In  case  of  improper  use  or  modifications,  repair  activities  by  unauthorised  staff  as  well  as  improper 
handling any warranty and liability on the part of the manufacturer ceases to apply. 
 
The device has to be returned to the supplier in its original packaging.   
 

Retain the original packaging for such transport. 

 
 

Please note:

 Please consider the legal provisions regarding disposal of the packaging, accessories or, 

at a later date, of the device itself.  

 
 
 
 

Содержание PDV30

Страница 1: ...PODIAVAC PDV30 Operating Instructions ...

Страница 2: ... the rotational direction of the tool 6 Switching the hand piece 6 Tool change 7 Separate switching on and off of the suction system 7 Switching off the device 7 General notes 7 General care maintenance and repair 8 Cleaning the device 8 Replacing the dust bag 8 Cleaning the hand piece 8 Disinfection cleaning and storing the tools 9 Maintenance 9 Malfunctions and trouble shooting 10 Changing the f...

Страница 3: ...aged during transport If there is any doubt you should immediately contact our customer service Please read the following safety instructions carefully Adequate ventilation of the device is necessary to prevent inside the unit too much heat Imagine the PODIAVAC should not be placed in the vicinity of heat sources such as e g radiators or air vents The ventilation slots on the device must not be bl...

Страница 4: ... of the tool ROTATION 8 Counter clockwise clockwise rotation of the tool L R 9 Service LED for maintenance 10 Memory for stored speed values 2x 11 Speed control rotation Increase speed Reduce speed 12 Display for vacuum power 13 ON OFF pushbutton for suction Vacuum 14 Level indicator dust bag 15 Control of suction capacity Increase capacity Reduce capacity 16 Memory for stored vacuum values 2x 3 2...

Страница 5: ...evice below the fixing hole for the hand piece Mains supply Connect the supplied power cable with the power inlet on the right side of the device Attention Check the voltage V specified on the label before connecting the device to the supply voltage Plug the power cord into a wall outlet with adequate protection contact The device is now ready for operation Warranty The term of the warranty for Po...

Страница 6: ...utton of the speed controller 11 to increase or reduce the speed The speed is controllable in steps of 1000 from 5000 up to 12000 and Steps of 2000 from 12000 up to 30000 min 1 The selected speed is displayed by the corresponding display 6 on the front panel Setting of suction Press the or pushbutton of the vacuum power control 15 to increase or reduce suction The suction capacity vacuum set is di...

Страница 7: ...hed on again at this point Separate switching on and off of the suction system Press the ON OFF pushbutton VACUUM 13 for the suction system The suction system is switched off The grinder continues to operate Press the ON OFF pushbutton VACUUM 13 for the suction system once again in order to switch the suction system once again Switching off the device Press the I O pushbutton 5 The device is now i...

Страница 8: ...r damaging it The guide pins of the connection of the hand piece have to be inserted into the guiding holes of the suction chamber Make sure that the contact pin is not damaged by any jamming which may occur Cleaning the hand piece NEVER clean the hand piece using boiling water chemicals wire brushes or ultra sound NEVER immerse the hand piece in a disinfectant or ultrasonic bath NEVER place or im...

Страница 9: ...nd repair may only be carried out by authorised skilled staff of the manufacturer Unauthorised persons are not permitted to open the device under any circumstances In case of changes and repairs by staff not authorised to that end any warranty or liability on the part of the manufacturer ceases to apply During the regular maintenance of the device an examination according to DIN VDE 0702 0751 has ...

Страница 10: ...Grinder is blocked Check connection between the plug of the hand piece and the device Ensure unobstructed operation of the grinder Start Stop pushbutton 7 Reduced suction capacity Dust bag is full Exchange the dust bag In case of further problems or malfunctions please contact your dealer or the manufacturer Changing the fuse Caution Fuses may only be exchanged after the power plug has been unplug...

Страница 11: ...vice incl the hand piece to our customer service department for maintenance If required you can request a replacement device In order to be able to continue working switch off the device with the Start Stop pushbutton 7 and switch on once again As long as no device maintenance happened this procedure must be repeated at every switching on the device Please remember If the maintenance intervals are...

Страница 12: ...NSK Wellness Technology GmbH Address Portitzer Strasse 69 d D 04425 Taucha Germany We certifies that the product Purpose of equipment Foot Care Equipment Type of equipment PODIAVAC Article no 9160 1 in accordance with the following standards EN 61000 6 1 2001 EN 61000 6 3 2001 Taucha 2012 01 01 NSK Wellness Technology GmbH Managing Director Dieter Bauch Operating instructions GA9160_4 E_V2_0 2012 ...