20
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
IEC 60601
test level
2 kV for power
supply lines
6 kV contact
8 kV air
± 1 kV line(s)
to line(s)
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
3A / m
3A / m
Compliance
level
6 kV contact
8 kV air
2kV for power
supply lines
± 1 kV line(s)
to line(s)
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Electromagnetic environment -
guidance
The Medical LED Headlight is intended for use in the electromagnetic environment specified
below. The customer or the user of The Medical LED Headlight should assure that it is used in
such an environment.
Floors should be wood, concrete or
ceramic tile. If floor are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of The
Medical LED Headlight requires
continued operation during power
mains interruptions, it is
recommended that The Medical
LED Headlight be powered from an
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Guidance and manufacture’s declaration – electromagnetic immunity
Guidance and manufacture’s declaration – electromagnetic immunity –
for all ME EQUIPMENT and ME SYSTEMS
NOTE:U
T
is the a.c. mains voltage prior to application of the test level.