3. Check the Pommel locks in place and does not
slide out when pulled (see section 4.4, 5.7).
4. Check for any damage to the soft moulded parts
(seat liners, headrest, armrests, pommel). If any
of these parts are damaged, they should be
replaced.
5.
Raise the plastic cover on the gas strut (pulling
upwards) and clean the lower part of the gas
strut with water. Lower the plastic cover and
push down to secure.
7
7
6
6. Check the Tilt-in-Space mechanism for smooth
operation. Make sure it locks fully when the red
push buttons are released (see section 5.5).
Check the assembly for hydraulic
fl
uid leaks.
Stop using the chair immediately if the
mechanism becomes dif
fi
cult to operate, fails to
lock correctly or leaks hydraulic
fl
uid.
7. Check the security of all adjustment knobs
and fasteners. Pay particular attention to the
armrests (see section 4.7), headrests (see
section 5.4) and anterior supports (see section
6.2).
8. Check all straps and harnesses for any signs
of damage to the webbing or to the buckles.
Damaged items must be replaced immediately.
Please contact NRS Healthcare Customer Services
to report a fault or book an inspection by a service
technician.
Refer to section 13.0 for a detailed list of spare parts.
7.2 Annual Inspection
We recommend this product is checked annually by
a quali
fi
ed technician for safe operation and repair of
any damaged or missing parts.
7.3 Cleaning & Disinfection
The chair should be cleaned at regular intervals,and
dried fully after each use.
Cleaning the chair is an ideal opportunity to identify
any signs of wear or damage (see section 7.0).
Regular cleaning is also essential to minimise
risk of infection transmission between users and
carers. This is particularly important in a multi-user
environment.
Particular attention must be taken to clean the
commode pan / chair interface, between the liners,
pommel, seat moulding and gas strut.
WARNING:
There is a risk of infection
transmission to the carer when cleaning the
chair. Protective equipment should be worn
during any cleaning or maintenance
•
Clean regularly using a general purpose, neutral
detergent and warm water.
• Please check the suitability of the cleaning
product before use. Do not use abrasive cleaners
or cloths as these could permanently damage
the surface and invalidate any warranty.
•
To disinfect, wipe with a dilute bleach solution,
and rinse thoroughly with fresh water.
•
Allow product to fully dry prior to use/storage.
•
Check the chair for functionality after cleaning
the chair
7.4 Lap-strap and Harness Cleaning
•
Clean with sponge and warm soapy water
•
Do not use bleach or solvents
•
Do not steam clean, tumble dry or dry clean.
8.0 PRODUCT DISPOSAL
This product should be thoroughly cleaned prior to
disposal.
This product is manufactured from metal and plastic
and can be disposed of through Local Authority
Household Waste and Recycling Centres or through
Local Trade Waste Collectors.
9.0 PRODUCT GUARANTEE
NRS Healthcare guarantee this product for a period
of 12 months, from the date of purchase, against
faulty materials and workmanship.
This guarantee does not cover; abuse, misuse,
interference or tampering with any part of the product.
Any guarantee issued by NRS Healthcare does not
affect your statutory rights.
10.0 PRODUCT
STANDARDS
BS EN 12182: 2012 (Assistive products for persons
with disability. General requirements and test
methods)
BS EN ISO 14971:2012 (Application of risk
management to medical devices)