MTC-20/2S User Manual
_______________________________________________________________________________________________________________
In Figure 5 the time course of temperature regulation of the MTC-20/2S is shown. As
resistive load a bulb (12V, 21W) was used. The temperature sensor was fixed directly on the
bulb and sensor and bulb were wrapped with foam. Starting temperature was 27,3 °C. Starting
conditions were: LIMITER set to 0, GAIN set to 0, INTEGRATION off.
25
27
29
31
33
35
37
39
0
100
200
300
400
500
600
700
800
900
1000
time (s)
temperature (°C)
1
2
3
Figure 5: Time course of temperature regulation of the MTC-20/2S.
1
: LIMITER was set to 100%, GAIN was set to a low level
2
: GAIN was set to a slightly higher value (about 10% of the whole scale)
3
: INTEGRATION was turned on at a low level
___________________________________________________________________________
version 2.7
page 15