ISO-STIM 01B User Manual 

version 1.8 

 

 

page 9 

3.5. Operation 

3.5.1.

 

Connections – Noise Reduction 

 
The ISO-STIM 01B comes with a power supply with SubD cable, which is connected to the 
stimulus isolation device. The corresponding connector at the ISO-STIM 01B is located at the 
rear panel. 
 

Note

: A possible source of noise is the chassis of the stimulator. In a noisy environment, it may 

act as an antenna picking up 50 Hz or 60 Hz noise respectively, especially when operating with 
current stimuli. In this case, the noise is substantially reduced if the chassis is connected to 
ground (see also #

11

, Figure 1). 

 
The stimulation electrode is connected to the isolated output (see Figure 1). We recommend to 
use a bipolar electrode for stimulating nerve bundles. This will produce only a short stimulation 
artifact at the beginning. 
 

Remember

: If unipolar electrodes are used the whole slice will be stimulated resulting in a DC 

shift of the whole slice. 
 

3.5.2.

 

Operation Modes 

 

BIPOLAR: 

 
The isolator is capable to generate biphasic stimuli by switching the stimulus mode switch to 
BIPOLAR ON. In this mode, the stimulus that is set by the user will be followed by a second 
stimulus with the same amplitude and duration, but with reverse polarity. There is no delay 
between first and second part of the stimulus (except IPG is active, see below). For instance, if 
the user sets a positive stimulus of +1 V amplitude and 100 µs duration, this stimulus will be 
followed immediately by a negative stimulus of –1 V and 100 µs duration. 
 

BIPOLAR with IPG: 

In bipolar mode, a delay between the first and the second, inverse stimulus can be inserted, the 
interphase gap (IPG, see Figure 4). IPG time is set by (

#16

 

 

Figure 4: Interphase gap (IPG) 

 

OPERATION IN THE 10 mA (max.) RANGE 

 
According to Ohm’s law, with 10 V voltage 10 mA current can be achieved 

only

 if the 

resistance of the stimulus electrode is not higher than 1 k

. Otherwise, the current will be less. 

Содержание ISO-STIM 01B

Страница 1: ...STEM DESCRIPTION FOR THE ISO STIM 01B STIMULUS ISOLATOR VERSION 1 8 npi 2016 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support npielectronic com http www npielectronic com ...

Страница 2: ...e for High Voltage Instruments 4 3 ISO STIM 01 B 5 3 1 System Description 5 3 2 Description of the Front Panel 5 3 3 Description of the Rear Panel 8 3 4 ISO CAB optional accessory 8 3 5 Operation 9 3 5 1 Connections Noise Reduction 9 3 5 2 Operation Modes 9 4 Power Supply 10 5 Technical Data ISO STIM 01B 11 ...

Страница 3: ...ge before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Before opening the cabinet unplug the instrument Unplug the instrument when replacing the fuse or changing line voltage Replace fuse only with an appropriate specified type 3 STATIC ELECTRICITY Electronic equipment is sensitive to static d...

Страница 4: ...er supplies to protective earth 2 Do not touch connections unless the instrument is turned off and the capacitance of both the load and power supply are earthed 3 Allow adequate time for discharge of internal capacitance of the power supply 4 Do not ground yourself or work under wet or damp conditions 5 Servicing should be only done by qualified personnel aware of the hazards 6 If in doubt return ...

Страница 5: ...T INPUT allows the pulse length to be externally controlled whereas amplitude and polarity are controlled by the device itself For both timing modes the generated pulses can be unipolar or bipolar In bipolar mode an interphase gap IPG of adjustable length can be inserted between the two pulses 3 2 Description of the Front Panel Figure 1 Front panel view of ISO STIM 01B In the following description...

Страница 6: ...MODE the limit 100 V is set by the corresponding voltage at the electrode due to Ohm s Law V I R i e this control indicates during an experiment in CURRENT OUTPUT MODE that the electrode resistance increases and the amplifier is not be able to provide the current which is set if the electrode resistance increases further 8 STIM LED LED indicating the application of a voltage or current stimulus 9 ...

Страница 7: ...nge to be set with the AMPLITUDE potentiometer 5 16 IPG TIME µs potentiometer with switch Potentiometer to select the interphase gap time IPG in bipolar stimulation mode see 3 5 2 IPG is turned off in the leftmost position switch 17 DIRECT INPUT TTL BNC connector A TTL signal can be fed in here to trigger the output of the ISO STIM 01D Amplitude will be set by 5 6 15 In unipolar mode 4 the output ...

Страница 8: ...ector are connected to the ISOLATED OUTPUT connectors on the front panel of the ISO STIM 01 D Each of these stimulation lines has a shield which is connected to the common green SHIELD connector To reduce noise during recordings this connector can be connected to the CHASSIS connector at the ISO STIM 01 D front panel The other end of the cable provides open wires to which different connector types...

Страница 9: ...ed the whole slice will be stimulated resulting in a DC shift of the whole slice 3 5 2 Operation Modes BIPOLAR The isolator is capable to generate biphasic stimuli by switching the stimulus mode switch to BIPOLAR ON In this mode the stimulus that is set by the user will be followed by a second stimulus with the same amplitude and duration but with reverse polarity There is no delay between first a...

Страница 10: ...left and rear panel right view of the ISO STIM 01B power supply The power supply is connected to mains with a standard power cord It has a POWER ON switch on the rear panel Important Only use the supplied power supply to power the ISO STIM 01B Using different power supplies might result in damage of ISO STIM 01B and or power supply Important This power supply is designed for use with the ISO STIM ...

Страница 11: ...bipolar selected by toggle switch Output voltage 100 V max isolated Output current 10 mA max isolated IPG interphase gap 50 500 µs Power requirements 100 240 V AC only use supplied power supply Fuses Power Supply 1 25 A slow Dimensions ISO STIM 01B 246 x 188 x 89 mm W x D x H Power Supply 103 x 169 x 59 mm W x D x H ...

Отзывы: