2. EPMS-07 Modular Plug-In System
2.1.
General System Description / Operation
The npi EPMS-07 is a modular system for processing of bioelectrical signals in
electrophysiology. The system is housed in a 19” rackmount cabinet (3U) has room for up to
7 plug-in units. The plug-in units are connected to power by a bus at the rear panel.
The plug-in units must be kept in position by four screws (M 2,5 x 10). The screws are
important not only for mechanical stability but also for proper electrical connection to the
system housing. Free area must be protected with covers.
2.2. EPMS-07
Housing
The following items are shipped with the EPMS-07 housing:
EPMS-07 cabinet with built-in power supply
Mains cord
Fuse 2 A / 1 A, slow
Front covers
In order to avoid induction of electromagnetic noise the power supply unit, the power switch
and the fuse are located at the rear of the housing.
2.3. EPMS-E-07
Housing
The following items are shipped with the EPMS-E-07 housing:
EPMS-E-07 cabinet
External Power supply PWR-03D
Power cord (PWR-03D to EPMS-E-07)
Mains chord
Fuse 1.6 A / 0.8 A, slow
Front covers
The EPMS-E-07 housing is designed for low-noise operation, especially for extracellular and
multi channel amplifiers with plugged in filters. It operates with an external power supply to
minimize distortions of the signals caused by the power supply.
2.4. PWR-03D
The external power supply PWR-03D is capable of driving up to 3 EPMS-E housings. Each
housing is connected by a 6-pole cable from the one of the three connectors on the front panel
of the PWR-03D to the rear panel of the respective EPMS-E housing. (see Figure 1, Figure 3).
A POWER LED indicates that the PWR-03D is powered on (see Figure 1). Power switch,
voltage selector and fuse are located at the rear panel (see Figure 2).
Note
: The chassis of the PWR-03D is connected to protective earth, and it provides protective
earth to the EPMS-E housing if connected.