background image

DPA-2FX User Manual 

___________________________________________________________________________ 

___________________________________________________________________________ 
version 1.8 

 

 

page 10 

(

13

) GAIN ZERO ADJ. trimpots 

 
Trimpots to compensate for amplifier offsets. If the baseline is not zero even if the 
input is grounded, this offset can be compensated independently for the two GAIN 
stages. Compensation procedure: 

 

 

Set the input coupling switches to zero to ground the inputs, and set the HIGH PASS 
FILTER to 400 Hz. 

 

Set the GAIN range switch to 1, the GAIN factor switch to one and adjust the baseline to 
zero using the left trimpot. 

 

Switch the GAIN factor switch to all positions and readjust to zero if necessary. 

 

Set the GAIN range switch to 10. If an offset is present, compensate it using the right 
trimpot. 

 

Set the GAIN range switch to 100 and readjust the baseline if necessary. 

 

4. Literature 

 

 

Barmashenko, G., Eysel, U. T., & Mittmann, T. (2003). Changes in intracellular calcium 
transients and LTP in the surround of visual cortex lesions in rats. 

Brain Res.

 

990

, 120-

128. 

 

Boulton, A. A., Baker, G. B. & Vanderwolf, C. H. (eds.) (1990). 

Neurophysiological 

Techniques, Basic Methods and Concepts

. Humana Press, Clifton, New Jersey. 

 

Huemmeke, M., Eysel, U. T., & Mittmann, T. (2002). Metabotropic glutamate receptors 
mediate expression of LTP in slices of rat visual cortex. 

Eur.J.Neurosci.

 

15

, 1641-1645. 

 

Huemmeke, M., Eysel, U. T., & Mittmann, T. (2004). Lesion-induced enhancement of 
LTP in rat visual cortex is mediated by NMDA receptors containing the NR2B subunit. 

Physiol

 

559

 , 875-882. 

 

Schulz, D., Huston, J. P., Jezek, K., Haas, H. L., Roth-Harer, A., Selbach, O., & Luhmann, 
H. J. (2002). Water maze performance, exploratory activity, inhibitory avoidance and 
hippocampal plasticity in aged superior and inferior learners. 

Eur.J.Neurosci.

 

16

, 2175-

2185. 

 

Kettenmann, H. & Grantyn, R. (eds.) (1992). 

Practical Electrophysiological Methods

Wiley-Liss, New York 

 

Lalley, P.M., A.K. Moschovakis and U. Windhorst (1999) Electrical Activity of Individual 
Neurons In Situ: Extra- and Intracellular Recording, in: U. Windhorst and H. Johansson 
(eds.) Modern Techniques in Neuroscience Research, Springer, Berlin, New York 

 

Müller, Ch.M. (1992) Extra- and Intracellular Recording in the Slice, in: Kettenmann, H. 
& Grantyn, R. (eds.) Practical Electro-physiological Methods, Wiley-Liss, New York 

 

Ogden, D. (ed.) (1992) Microelectrode Techniques - The Plymouth Workshop Handbook, 
Second Edition, The Company of Biologists Ltd., Cambridge 

 

Seidenbecher, T. and H.C. Pape (2001) Contribution of intralaminar thalamic nuclei to 
spike-and-wave-discharges during spontaneous seizures in a genetic rat model of absence 
epilepsy, European Journal of Neuroscience, Vol. 13:1537-1546 

 

Windhorst, U. and H. Johansson (eds.) (1999) Modern Techniques in Neuroscience 
Research, Springer, Berlin, Heidelberg, New York 

Содержание DPA-2FX

Страница 1: ...FOR THE DPA 2FX DIFFERENTIAL AMPLIFIER FILTER MODULE FOR EPMS SYSTEMS VERSION 1 8 npi 2014 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 support...

Страница 2: ...PMS 07 Modular Plug In System 4 2 1 General System Description Operation 4 2 2 EPMS 07 Housing 4 2 3 EPMS E 07 Housing 4 2 4 PWR 03D 4 2 5 System Grounding 5 EPMS 07 5 EPMS E 07 5 2 6 Technical Data 5...

Страница 3: ...only at 115 230 Volt 60 50 Hz AC Please check for appropriate line voltage before connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact conn...

Страница 4: ...n of electromagnetic noise the power supply unit the power switch and the fuse are located at the rear of the housing 2 3 EPMS E 07 Housing The following items are shipped with the EPMS E 07 housing E...

Страница 5: ...ound plugs on the rear panel of the instrument It is not possible to predict whether measurements will be less or more noisy with the internal ground and mains ground connected We recommend that you t...

Страница 6: ...t are BNC connectors The input impedance is 10 M Thus the module can be used for direct recording of bioelectric signals e g from peripheral nerves Using the differential input configuration common mo...

Страница 7: ...__________________________________________________________ ___________________________________________________________________________ version 1 8 page 7 3 4 Description of the Front Panel and Operati...

Страница 8: ...ge switch 3 position switch for selecting the OFFSET range 0 1 V 1 V or 10 V 2 OFFSET potentiometer 10 turn potentiometer for compensating for a DC OFFSET Note The overall OFFSET compensation is the p...

Страница 9: ...lter with bandpass characteristic can be implemented easily 6 OVER OVER LEDs LEDs indicating that the DPA 2FX exceeds it s linear range 10 V 7 OUT connector BNC connector providing the conditioned sig...

Страница 10: ...ptors mediate expression of LTP in slices of rat visual cortex Eur J Neurosci 15 1641 1645 Huemmeke M Eysel U T Mittmann T 2004 Lesion induced enhancement of LTP in rat visual cortex is mediated by NM...

Страница 11: ...0 1 1 V or 10 V set by toggle switch HIGHPASS Filter single pole attenuation 6 dB octave corner frequencies Hz 3 30 100 300 400 500 600 800 1k 3k 10k LOWPASS Filter single pole attenuation 6 dB octave...