The TagTemp-NFC is an electronic temperature logger. The sensor 

measures this physical quantity and the values are stored in memory. 

These data are collected by a smartphone equipped with an NFC 

interface (not included). 
The  LogChart-NFC  software  enables to set up the equipment 

operation mode. It is also used for viewing acquisitions. Other 

parameters such as end of measurements, logging interval, etc., are 

easily defined through the LogChart-NFC software. 
The LogChart-NFC application is targeted to ANDROID™ platform 

and must be downloaded from Google Play Store

2

 

MECHANICAL INSTALATION 

The TagTemp-NFC can be fixed in two ways: by using of screws or 

double-sided tape. 
For mounting with screws, you should detach the front film of the 

equipment carefully, exposing two through holes where TagTemp-

NFC may be screwed, as shown in Fig.1. They should be used M3 

flathead screws. When finished, you should stick extra film that 

comes with the product to cover again the front of the device. 
The other way is to use a double-sided adhesive tape on the back of 

the equipment, avoiding placing it above the identification label. 
In both cases, the temperature sensor should be placed near the 

place where you want to measure the temperature. 

 

Fig. 01 - Mounting holes on the TagTemp-NFC 

 

 
 

3

 

TECHNICAL DATA 

Measure Range 

Temperature: -40.0 °C to 70.0 °C. 

Accuracy 

±

 0.5 ºC @ 25 ºC. 

±

  1 °C max. over the full measuring 

range. 
Note: A measurement error can be 

ruled out using the OFFSET parameter 

in the LogChart-NFC software. 

Measurement 

Resolutions 

Temperature: 0,1 °C / 0,1 ° F. 

Memory capacity 

Up to 4020 logs. 

Measurement Interval 

5 minutes min. 
2 hours max. 

Supply 

2 x 3.0 V lithium battery, built-in, non-

replaceable. 

Estimated battery 

autonomy 

 

Above 300 days – 5-minute logging 

interval. 

 

Above 400 days – 30-minute 

logging interval. 

Operating temperature  From -10 °C to 70 °C 
Dimensions (L x W x H)   Main body: 65 x 44 x 8 mm. 
RF Interface 

ISO 15693 (NFC-V) compatible. 

Logger-Smartphone 

data transfer time 

Approx. 20 seconds (depends on the 

smartphone processing speed). 

LogChart-NFC 

software 

Android version: 4.0 or above. 
Languages: Portuguese and English. 

PC Interface 

USB to NFC converter interface 

(optional), compatible with LogChart-II 

for Windows ®. 

Material 

Polyamide  

Protection 

All electronics are protected by the 

housing. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание 8813700000

Страница 1: ...he other way is to use a double sided adhesive tape on the back of the equipment avoiding placing it above the identification label In both cases the temperature sensor should be placed near the place where you want to measure the temperature Fig 01 Mounting holes on the TagTemp NFC 3 TECHNICAL DATA Measure Range Temperature 40 0 C to 70 0 C Accuracy 0 5 ºC 25 ºC 1 C max over the full measuring ra...

Страница 2: ...lowed by a slow one beginning or ending of records If the LED is always off there are two possible reasons 1 The equipment is in low power mode 2 The battery is dead The LED flashing light is intentionally weak and can be difficult to be viewed in bright places 4 4 Button The S button on the front panel can be used to start and or stop the records depending on the configuration chosen To start the...

Страница 3: ...takes a few seconds depending on the smartphone model The application has six screens 1 Status 2 Setup 3 Events 4 Data Visualization 5 Preferences 6 Manage Collects 5 2 1 Device Status Fig 04 Status Screen Last Recorded Value Displays the last recorded value performed by the TagTemp NFC Not necessarily the actual temperature Loggings Displays the basic information of the TagTemp NFC Log Interval D...

Страница 4: ... digital input Digital Input Events Number of events occurred Last Event Indicates the time of the last event occurred Last State Indicates the state of the last event occurred Longest Period in ON Indicates the longest time that remained in ON state High Excursions Indicates when the measured value exceeds the upper limit specified First Occurrence Time and value of the first occurrence above the...

Страница 5: ...nce as shown in Fig 10 Fig 11 and Fig 12 If there is no occurrences in the Digital input neither high or low Excursions messages appear in the Fig 13 Fig 14 and Fig 15 Fig 10 Digital Input Events Screen Fig 11 High Excursions Screen Fig 12 Low Excursions Screen Fig 13 Message for Events in the Digital Input screen for No event Fig 14 Message for High Excursions for No occurrences Fig 15 Message fo...

Страница 6: ...ences screen Fig 18 Share Screen 5 2 5 Preference Settings Default share options The user can set the preferences screen according to the desired data received through sharing Keep downloaded data in smartphone Maintains files previously selected in the sharing options in a folder named LogChartNFCFiles in the internal memory of the smartphone Export Logs CSV Share a file in CSV format with all re...

Страница 7: ...equipment In order to configure the device is necessary to place the TagTemp NFC over the NFC USB reader After the serial port is selected click on the button The Parameters Configuration screen is displayed In this screen the user can define the equipment operation mode and also obtains general information about the device Fig 23 Parameters Configuration screen Fields are 1 Title In this field th...

Страница 8: ...zation Data collected are send from equipment to PC Data can be collected any time at the end of the acquisition process or while they are being acquired If data collection takes place during the acquisition process the process will not be interrupted following the equipment configuration It is important to wake up the equipment before starting communication 6 5 Offloading data Data offload is acc...

Страница 9: ... displayed by pressing the icon Parameter Viewing Fig 27 General information 6 7 Exporting data Logged data can be exported to different file formats for further analysis To export access File Export or click the icon 6 8 Diagnostic Screen Battery Level Indicates the level of battery voltage When you observe that the battery level is low replace it immediately for proper device operation Device St...

Отзывы: