RDC.OD.TP-2020.000 OM
Operation manual
Coating Thickness Gauge NOVOTEST TP-2020
Page 21
6.
To turn the device off, press and hold «
».
2.3.13 Measurement of the air temperature, humidity, and dew point
The air temperature, air humidity, and dew point should be measured with the STMD probe.
To measure the air temperature, humidity, and dew point, do the following steps:
1.
Connect the STMD probe to the device (c. 2.3.3).
2.
Holding the probe by the handle (where parameters should be measured), make measurement
before turning the device on and holding «
» button until the screensaver appears.
3.
After turning on, the device will start measurements automatically.
4.
The display will show the measurement results (fig. 2.29).
Figure 2.29 – The graphic display when using the STMD probe
5.
To turn the device off, press and hold «
».
2.4 Connection of the device to PC
2.4.1 Software installation
Connecting the device to a PC, user can transfer the information stored on the memory card to a PC.
To connect the device to a PC:
1.
Copy the driver «CP210x_VCP_Windows» and the program «AWP TP1M» (archives of .zip
and .rar format) to the hard drive of the PC or download the current versions from the official
website.
2.
After downloading, unzip the data, and two folders will be obtained: driver and program (fig.
2.30).
Figure 2.30 – The downloaded software
3.
Install the driver to PC.
Note – After notice about successful driver installation, it is advisable to restart a PC.
4.
Now, user can connect the device to a PC by using the USB cable from the delivery set. After
connection, the computer will detect the new device and install the driver to use it.
5.
Install the program to use the device, to do that, launch the program installation file (setup.exe)
from the folder «AWP TP1M», and press «INSTALL» (fig. 2.31).
Air humidity
Air temperature
Dew point