RDC.OD.TP-2020.000 OM
Operation manual
Coating Thickness Gauge NOVOTEST TP-2020
Page 16
1.
Figure 2.14 – Calibration of the first value
If the obvious error appears after measurement, press «
» and select «RESET», and the series
of measurements will be removed.
Then, do the same actions with other blocks (fig. 2.15).
Figure 2.15 – Calibration by three reference blocks
To save the calibration, press «back» and confirm saving by pressing «
».
2.3.5 Working in «ARCHIVE» mode
When the «Archive» mode is selected, the device goes to viewing of the saved measurements (fig.
2.16). The device memory can save up to 1024 measurements.
Figure 2.16 – Viewing of the saved measurements
2.3.6 Working in «SETTINGS» mode
When the «SETTINGS» item is selected, the device goes to settings of the following parameters
(fig. 2.17):
−
«LANGUAGE»: selection of the language of the device menu (English, Russia, and others
are available);
−
«BRIGHTNESS»: changing of the display brightness (from 0 to 100%, discreteness 10%);
−
«CONTRAST»: changing the display contrast (from 0 to 100%, discreteness 10%);
−
«AUTO OFF»: setting of the automatic shutoff when the device is not used (OFF, 1 min,
5 min, 10 min, 30 min);
−
«SOUND»: setting of the sound signal when pressing of buttons and measurement
(Key +Meas., Meas., Key, OFF);
−
«RETRO»: allows returning to the measurement mode with the latest measurements after
the device rebooting;
−
«SCALE»: selection of a measuring unit (mm, mkm, inch, mil);
−
«MAT. BASE»: selection of the substrate of the test object (F – ferromagnetic base, NF –
non-ferromagnetic base, B1, B2);
−
«MAT. COVER»: selection of the coating material of the test object (DI – dielectric coating,
and user coatings: U1, U2, U3, U4);
Current code value
The obtained value of the
function (polynomic)
Measurement number