50
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output lines
±2 kV
N/A
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line to line
±2 kV line to earth
±1 kV line to line
±2 kV line to earth
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
Mains power quality
should be that of a
typical commercial or
hospital
environment.
Power frequency
(50/60 Hz)
magnetic field
3 A/m
3 A/m
Power frequency
magnetic fields
should be at
levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level = 120V and 230V
Normal operation: The Optune Treatment Kit is working properly when the blue LED
surrounding the TTFields button are lit and no notification signal sounds. The ICH9100
charger is working properly when all the LEDs are lit. The SPS9100 power supply is
working properly when the blue LEDs surrounding the TTFields button on Optune
Treatment Kit are lit and no notification signal sounds.
Содержание Optune NovoTTF 200A
Страница 1: ...USER MANUAL...
Страница 2: ...Document number QSD EUUM 002 EN Version 02...
Страница 19: ...19...
Страница 25: ......
Страница 29: ...21 Full Charge Empty 25 Charge 50 Charge 75 Charge...
Страница 41: ...33 Serial Number Class II equipment per IEC 60601 1 Medical device...
Страница 53: ...45 BS EN 62368 1 Audio video information and communication technology equipment Safety requirements...