background image

 

3.Calibration  

The three Quantadats ,with matched Nsens probes ,are supplied calibrated with a traceable 
certificate and no calibration checks should be necessary under normal circumstances, only if 
readings are supect for some reason. A calibration kit ( SC humidity pots) is suplied only if this 
is the case.  

If calibration is needed: 

Calibration may only be carried out as a check , no adjustment is possible with this kit. 
Calibration may only be checked in RELATIVE HUMIDITY MODE so you need to set each 
Quantadat to read Rh% ( see Step 3). Enter menu, select Measuring Point A and select 
Relative Humidity. Rememebr to revert back to g/Kg ( mixing ratio) after any calibration 
checks. 

The Nsens priobes are supplied pre-calibrated and rady for use. All of the calibration 
adjustment settings are stored inside the memory of each Nsens probe. However, it is 
essential to have the means to check calibration with the SC calibration salts set at any teime 
to ensure performance of teh system. The calibration tolerance to expect is + / - 2% relative 
humdity and therfore the maximum difference you should see between all three probes should 
be 4% relative humdity. 

Using the 

SC

 (Sensor Check) calibration salts  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Novasina calibration service 

Storage and temperature stability: 

Ensure that the calibration salts 

SC

 have been kept inside the sealed 

containers and they have evidence of salt crystal and moisture inside. 
Shake them up ready for use. Ensure that the 

SC 

calibration salts are 

at the same room temperature as the humidity probes you are about 
to calibrate.  

Use only between 

15 

– 30

 Celcius. 

Application to probe: 

Remove the grey plastic cap from the 

SC

 calibration salt and shake it 

once more. Fit it to the probe you are calibrating and ensure there is a 
reasonably  air  tight  seal  around  it  (you  may  need  to  use  the  grey 
adaptor ring supplied with the instrument  if the probe is smaller  than 
the 18mm diameter of the inside of  the 

SC. 

Readings: 

Leave the 

SC

 calibration salt to equilibrate with the humidity probe for 

minimum of 30 minutes

 at stable temperature . 

Check the calibration (1% rh tolerance 

– see the humidity at different 

temperatures on the label on the side of the 

SC

).  

Examples of SC humidities  @ 25

 Celcius. 

SC33  ...................  32.8% 

SC58  ................... 57.6% 

SC84  ................... 84..3% 

Содержание Novasina Quantadat TRIPLEX

Страница 1: ...Novasina Quantadat TRIPLEX humidity system with nSens probes Set up commissioning manual...

Страница 2: ...gs can be interchanged without any issue Configuration may be required depending on the application eg the range of measurement the output parameter or the mA analogue output However the standard conf...

Страница 3: ...adat Transmitter dimmensions Nsense probe Green Brown White Power 24 V Analogue outputs 4 20mA CH2 20 to 80 C CH1 0 to 100 rh Power supply 24 V ac dc Output CH2 4 20mA Temp 20 to 80 C Output CH1 4 20m...

Страница 4: ...se instructions from the three front menu keys UP DOWN ENTER Set up of g Kg mixing ratio Scaling range of output in g Kg mixing ratio ENTER ENTER Measurement DOWN ENTER Analogue Outs DOWN ENTER Channe...

Страница 5: ...been recognized already this message appear You can choose Yes to continue and discard previous sensors Quantadat is now searching for connected sensors Step 2 Assign nSens to measuring points A to D...

Страница 6: ...parameters to channels 1 to 4 Analog Output equals display line 1 etc Enter Choose Source Measuring point A to D Choose measuring point as assigned in step 2 des Messpunktes A to D gem ss Schritt 2 If...

Страница 7: ...put range On Fail Output value setting in mA or V in case of a sensor failure e g interrupted communication with probe Adj Clear Deletion of adjustment This menu command is only shown if the analogue...

Страница 8: ...is 2 relative humdity and therfore the maximum difference you should see between all three probes should be 4 relative humdity Using the SC Sensor Check calibration salts Novasina calibration service...

Страница 9: ...pletion of installation Installation and comissioning is now complete and you are ready to put the Triplex Quantadats into operation For any further assistance please contact Novatron Scientific Ltd t...

Отзывы:

Похожие инструкции для Novasina Quantadat TRIPLEX