Step 1:  Connect nSens to Quantadat and start detection  

At first startup this message might appear:   

 

Choose YES to reach Sensor Setup directly 
 
 

 

Otherwise follow these instructions to detect sensors manually:  

Main Display:   

 

 

 

 

 
 

IF sensors have been recognized already this message 
appear. You can choose Yes to continue and discard 
previous sensors.  
 
 

 

Quantadat is now searching for connected sensors.  

 
 

 

Step 2: Assign nSens to 

measuring points 

A to D 

(Sensor Assign.) 

 

  

 

 

 

 
 

 

 

Press «Enter» and «arrow keys» to assign a 
serial number to a letter A to D.  Each serial 
number can only be assigned once.   
If just 1 sensor is connected only one letter 
is assigned. 
 

Press 

for Menu 

Press 

for Menu 

Scroll down 

  

Содержание Novasina Quantadat TRIPLEX

Страница 1: ...Novasina Quantadat TRIPLEX humidity system with nSens probes Set up commissioning manual...

Страница 2: ...gs can be interchanged without any issue Configuration may be required depending on the application eg the range of measurement the output parameter or the mA analogue output However the standard conf...

Страница 3: ...adat Transmitter dimmensions Nsense probe Green Brown White Power 24 V Analogue outputs 4 20mA CH2 20 to 80 C CH1 0 to 100 rh Power supply 24 V ac dc Output CH2 4 20mA Temp 20 to 80 C Output CH1 4 20m...

Страница 4: ...se instructions from the three front menu keys UP DOWN ENTER Set up of g Kg mixing ratio Scaling range of output in g Kg mixing ratio ENTER ENTER Measurement DOWN ENTER Analogue Outs DOWN ENTER Channe...

Страница 5: ...been recognized already this message appear You can choose Yes to continue and discard previous sensors Quantadat is now searching for connected sensors Step 2 Assign nSens to measuring points A to D...

Страница 6: ...parameters to channels 1 to 4 Analog Output equals display line 1 etc Enter Choose Source Measuring point A to D Choose measuring point as assigned in step 2 des Messpunktes A to D gem ss Schritt 2 If...

Страница 7: ...put range On Fail Output value setting in mA or V in case of a sensor failure e g interrupted communication with probe Adj Clear Deletion of adjustment This menu command is only shown if the analogue...

Страница 8: ...is 2 relative humdity and therfore the maximum difference you should see between all three probes should be 4 relative humdity Using the SC Sensor Check calibration salts Novasina calibration service...

Страница 9: ...pletion of installation Installation and comissioning is now complete and you are ready to put the Triplex Quantadats into operation For any further assistance please contact Novatron Scientific Ltd t...