NovaLynx Corporation_____________________________________________________________________________ 

 

 

200-WS-23 

Page 7 

Oct 2020 

 

6

 

CALIBRATION 

The 

200-WS-23 and 200-WS-23S Current Loop Wind Sensors 

are pre-calibrated and should require no 

adjustment over their lifetime. However, if the cable between the anemometer and the transmitter is 
extended beyond 250 feet (75 m), then resistance losses may introduce errors in the wind direction 
readings. 

A calibration procedure is beyond the scope of this manual. However, for those with the technical 
expertise, the board can be adjusted by means of multi-turn potentiometers located on the 4-20 mA 
driver circuit board. 

 

Offset 

Span 

Wind Speed 

R-33 

R-17 

Wind Direction 

R-26 

R-28 

 
 

The following table can be used as a guide if it becomes necessary to check the calibration of the 
monitoring equipment. Disconnect the 200-WS-23 or 200-WS-23S from the device before doing these 
tests. 
 

1.

 

Use a 4-20 mA signal generator to feed 6 mA into the selected channel on your monitoring 
device.  

2.

 

Compare the output of the monitoring equipment to the values in the table.  

3.

 

Switch the signal generator to output 18 mA. Compare the output to the table. 

4.

 

If both readings are higher or lower than in the table by the same amount, the zero offset 
parameter in the monitoring equipment needs to be adjusted. 

5.

 

If the readings are off by different amounts, then the slope parameter needs to be adjusted, 
and after that it may be necessary to adjust the zero offset also. 

 

 

mph

m/s

kph

degrees

compass

4.0

0.00

0.00

0.00

0

N

6.0

12.50

6.25

25.00

45

NE

8.0

25.00

12.50

50.00

90

E

10.0

37.50

18.75

75.00

135

SE

12.0

50.00

25.00

100.00

180

S

14.0

62.50

31.25

125.00

225

SW

16.0

75.00

37.50

150.00

270

W

18.0

87.50

43.75

175.00

315

NW

20.0

100.00

50.00

200.00

360

N

Wind Speed

Wind Direction

Loop Current

Содержание 200-WS-23

Страница 1: ...DOC 200 WS 23 UM 20201012 200 WS 23 User Manual Current Loop Wind Sensor 200 WS 23 Wind Speed Direction 200 WS 23S Wind Speed Only Phone 530 823 7185 Email nova novalynx com Website www novalynx com...

Страница 2: ...in the case of delicate or sensitive items If packing facilities are not available take the equipment to the nearest Post Office UPS or other freight service and obtain assistance with the packaging P...

Страница 3: ...CONTENTS 1 FORWARD 4 2 INTRODUCTION 4 3 SPECIFICATIONS 5 4 INSTALLATION 5 5 CONNECTIONS 6 6 CALIBRATION 7 APPENDIX A Wind Speed and Direction Wire Diagram 8 APPENDIX B Wind Speed Only Wire Diagram 9 2...

Страница 4: ...power supply is not needed The driver circuitry is protected in a NEMA enclosure suitable for indoor or outdoor mounting The anemometer is a 200 WS 01B Wind Speed sensor that includes 40 of cable and...

Страница 5: ...200 WS 23S KM 4 20 mA 0 200 kph Loop supply voltage range Connection Operating temperature range NEMA enclosure dimensions Cable to user equipment Measurement range Speed Constant Transducer type Spe...

Страница 6: ...Tighten the gland nut 3 Make sure there is a drip loop below the transmitter enclosure so that any moisture on the cable will not seep into the enclosure This is particularly important since the seal...

Страница 7: ...n of the monitoring equipment Disconnect the 200 WS 23 or 200 WS 23S from the device before doing these tests 1 Use a 4 20 mA signal generator to feed 6 mA into the selected channel on your monitoring...

Страница 8: ...2 3 4 GND WS REF REF EXC WD IMPORTANT CONNECT EARTH GROUND 200 WS 02F GREEN RED BLACK WHITE SHIELD BROWN 8 30 Vdc Wind Direction Signal 8 30 Vdc Wind Speed Signal 4 20 mA Input 8 30 Vdc 1 Signal Out...

Страница 9: ...ogger Power WIRING DIAGRAM 200 WS 23S JDC 9 03 2020 860 0029 01 1 2 3 4 GND WS REF REF EXC WD IMPORTANT CONNECT EARTH GROUND 200 WS 01B RED SHIELD BLACK 8 30 Vdc Wind Speed Signal 4 20 mA Input 8 30 V...

Отзывы: