53
3. Observe surface contact time.
• Ensure the Meter surface stays wet for the
recommended time (1 minute for the Clorox®
Germicidal Wipes or 2 minutes for the Super
Sani-Cloth® Germicidal Disposable Wipes) and
is allowed to air dry for an additional 1 minute.
NOTE:
If you must rewet the surface of the Meter,
use a new, fresh wipe.
4. Dispose of wipes
• Dispose of the used germicidal wipes in a
standard waste container.
5. Wash and sanitize hands.
• Wash your hands thoroughly with soap
and water.