If your Control Solution test results continue to fall outside the
range printed on the Control Solution vial label:
• The Nova Max Creat eGFR Meter and Test Strips may not be work-
ing properly.
• Do not use the Meter to test your blood.
• Contact Customer Service.
Testing a Control Solution
1. From the Home screen, press Run QC Test. The
Select QC screen displays. Select Run QC 1 or
Run QC 2.
2. Once the QC test is selected the Insert Test
Strip screen displays. Insert a test strip.
Running Nova Max Creat eGFR Control Solution