Pressure Connections

D/P Module Transmitter incorporates two (2) 1/8" NPT 
ports for HIGH and LOW differential pressure inputs.

 

vent plug for atmospheric reference (or low pressure  

 

input when measuring differential pressure)

 

connection for pressure input signal (high pressure  

 

input when measuring differential pressure)

When application is for psig (gage pressure), the low pressure 
port is not used. A vent plug should be installed in the low 
pressure port to allow transmitter to reference atmosphere. 
The low pressure port should 

never be plugged or sealed!

Signal Loop Connections

A plug-in type terminal connector is furnished for the 

+

 

Signal and 

 Signal inputs. Terminal block can be removed 

to make the cable connections and then re-inserted into 
the mating socket. Screw clamp terminals are provided 
for positive (

+

) and negative or common (

) sides of the 

signal loop.

As is typical of two-wire transmitters, all electrical power 
is supplied through the signal loop wiring. Refer to the 
following diagram for proper wiring of the 4–20 mA loop 
circuit:

E

S

1/8 in NPT

(Optional)

1/4 in Fittings

Low

Pressure

Ports

High

Pressure

Ports

4-20 mA

Output

Connection

+

-

+V

A

V

COM

R

L

KING-

GAGE

ZERO

SP

AN

+

-

D/P
MODULE

4-20 mA
Pressur

e

Transmitter

T
E
S
T

Plug-in Power/Signal

Connector

+   Ð

2

1

4-20 mA

Output

Connection

+

-

+V

A

V

COM

R

L

KING-

GAGE

ZERO

SP

AN

+

-

D/P
MODULE

4-20 mA
Pressur

e

Transmitter

T
E
S
T

Plug-in Power/Signal

Connector

+   Ð

2

1

Power Requirement

The D/P transmitter requires a source of DC power (14– 40 
Vdc) for operation. Minimum excitation voltage must be 
no less than 14 Vdc Any receiver installed on the signal 
loop (meters, data loggers, controllers, etc.) must be taken 
into account when determining the required power supply 
voltage to be used. The internal resistance of each device 
added together represents the total “load” residing on the 
signal loop circuit.

 

Load Capacity  =  (Supply Voltage 

12 Volts) 

 

  

 

  (ohms)   

 

.02

Load Capacity at Supplied Voltage

  

KING-GAGE

®

 tank processors and LevelBAR provide 24 Vdc 

excitation to power the signal loop circuit. (If the application 
requirement exceeds 700 ohms, an external power supply 
of appropriate voltage will be required.)

Signal Cabling

The 4–20 mA signal loop needs to be run using twisted pair 
(two conductor) cable. “Noise” or EMI (electromagnetic 
interference) does not generally create a problem since it 
is common to both wires in the pair and essentially cancels 
itself out. In most applications, non-shielded twisted pair 
instrumentation cable (20–22 awg) will be suitable for the 
signal loop between the LP2 and sensor/transmitter.

  

20 Vdc         24 Vdc        28 Vdc          32 Vdc           36 Vdc          40 Vdc  

400 ohms    600 ohms    800 ohms   1000 ohms    1200 ohms    1400 ohms

S

E

Содержание KING-GAGE 5900

Страница 1: ...e to Range Calculation Calibration Procedures Zero Span Adjustments for Differential Pressure Transmitters Model 5900 D P Module 4 mA to 20 mA Pressure Transmitter Multiple Tank Level Display KING GAGE A NOSHOK Company I 1010 West Bagley Road Berea Ohio 44017 I Ph 855 367 2494 I www king gage com ...

Страница 2: ...ties of fitnessforaparticularpurposeormerchantability aregiven bythisagreement Theexclusiveremedyfornonconformity of these goods shall be repair and or replacement of the nonconforming goods or parts Sellerwillnotbeliableforconsequentialdamagesresulting from breach of this agreement The term consequential damages shall include but shall not be limited to damage toallmachines equipmentandgoodsother...

Страница 3: ...pan Adjustment 6 Low Pressure Input for differential pressure 8 Calibration Test Setup Typical 9 Pressure Range Calculation 10 Troubleshooting 11 5900 1 2 Fittings 1 none 1 8 NPT connection 4 Swagelok 1 4 tube 1 8 NPT compatible with copper or poly tubing Range 0 0 5 psid 0 0 3 bar 1 0 10 psid 0 0 69 bar 2 0 15 psid 0 1 03 bar 3 0 30 psid 0 2 06 bar 4 0 50 psid 0 3 44 bar Model No Designations ...

Страница 4: ...r Supply Voltage 14 40 Vdc unregulated Power Supply Stability Effect on FSO less than 0 005 of span change in output per volt change at input terminals Burst Pressure 200 psi Span Adjustment Range Transmitter adjustment turndown is 3 1 for the 0 15 0 30 0 50 psid nominal ranges Nominal Minimum Range of Adjustment 0 5 psid 0 3 psid 0 83 in thru 0 138 in water 0 2 1 m thru 0 3 5 m water 0 10 psid 0 ...

Страница 5: ...Pressure Transm itter T E S T Plug in Power Signal Connector Ð 2 1 RL K IN G G A G E ZERO SPAN D P M O DULE 4 2 0 m A Pressure Transm itter T E S T Plug in Power Signal Connector Ð 2 1 Power Requirement The D P transmitter requires a source of DC power 14 40 Vdc for operation Minimum excitation voltage must be no less than 14 Vdc Any receiver installed on the signal loop meters data loggers contro...

Страница 6: ...t to the maximum depth hydrostatic pressure Measure the output using the meter terminal posts on the transmitter board 2 Adjust span potentiometer as necessary to increase output clockwise ordecreaseoutput counterclock wise until the signal displays as 20 00 or 20 01 mA on the test meter 3 Next step is to re check the zero output and slightly adjust if necessary In Process Zero Span Adjustment It ...

Страница 7: ...lease refer to the individual DIP switch positions listed for adjusting the existing setting of the transmitter The four 4 switches provide 16 increments of coarse adjustment Prior to changing any switch setting check the current output of the transmitter at the desired upper pressure value Also note the existing switch setting 1 Apply the desired maximum pressure to the sensing diaphragmoftheD Pt...

Страница 8: ...tter sensing element and the high pressure to the other side the internal tank pressure above the liquid low pressure is balanced across the sensing element This is also referred to as being equalized which is why we designate the low pressure connection as E CAUTION While the internal pressure above the liquid is not a factor in determining range span there may be instances where the sum of the l...

Страница 9: ...ipment MILLIAMMETER 3 1 2 digit minimum 0 05 accuracy 20 ohms max internal resistance DC POWER SUPPLY PRESSURE INDICATOR PSI or IN OF WATER 0 05 accuracy corresponding to transmitter range 5 psid nominal 0025 psi or 0693 of water 10 psid nominal 005 psi or 1385 of water 15 psid nominal 0075 psi or 2078 of water 30 psid nominal 015 psi or 4156 of water 50 psid nominal 025 psi or 6927 of water NOTE ...

Страница 10: ...rve represents the distance from the lowest point on the bottom of the tank to the installed sensor Full is the level of contents at which the tank is filled to desired level which is less than the actual full capacity of the tank The liquid contents in the tank has a specific gravity of 1 032 C 1 032 Specific Gravity B 300 Upper Level A 28 Reserve 300 28 1 032 27 6807 10 14 psid 300 28 1 032 281 ...

Страница 11: ...20 0 mA Check zero output of transmitter adjust zero to 4 00 mA if necessary Check span setting adjust to 20 00 mA at nominal range or calculated full pressure Maximum output less than 20 00 mA Check that applied pressure is within span limits for transmitter range Transmitter has 40 span adjustment Check span setting adjust to 20 00 mA at nominal range or calculated full pressure Higherresistivel...

Страница 12: ......

Отзывы: