5.5 How to log data to a file
1.
Select the correct operating mode of the instrument (see 5.1)
2.
Select [ Measure ]
3.
Select [ Continuous log ]
4.
Select the [ Interval ] – button to set the interval between each record measured. The number specifies
the interval between logging to the log file. By using the wheel, you may select 0-2 seconds interval in
0.1s steps, 2-10 second intervals in 1s steps, 10-60 second intervals in 10s steps or 60-600 second
intervals in 60s steps
5.
Select parameters for logging. Use the wheel or the up/down-buttons to select parameter, and then
select the [ Log ] – button to enable/disable logging of that parameter. For viewing the parameter live
on the screen while logging, use the [ Show ] –button
6.
The log files are named as follows:
<System>-<Frequency>-<Date>-<Serial number>.
Example: LLZ-110.300-2004-08-20-2
7.
Select [ Start ] to start logging to the log-file
8.
Select [ Stop ] to stop logging to the log-file
9.
Selecting [ Start ] again starts recording to another file with the following serial number according to
the naming convention described above