15
Microphone Cartridge Calibration Values
Microphone cartridges should be calibrated in accordance with their size and working principle. The table lists some
commonly used microphone cartridges and their corresponding calibration values applicable with the protection grid mounted.
Manufacturer
Microphone Type
Calibration Level
Norsonic
1220
113,85dB
1221
113,85dB
1223
113,85dB
1225
113,85dB
1226
113,85dB
1227
113,85dB
1228
113,85dB
1230
114,00dB
1233
114,00dB
1234
114,00dB
1235
114,00dB
1236
114,00dB
1237
114,00dB
1245
114,00dB
1246
114,00dB
Microtech
MK102
113,70dB
MK103
113,70dB
MK202
113,85dB
MK221
113,85dB
MK222
113,85dB
MK231
114,00dB
MK250
113,85dB
G.R.A.S.
40AC
113,85dB
40AD
114,00dB
40AE
113,85dB
Manufacturer
Microphone Type
Calibration Level
G.R.A.S. cont...
40AF
113,85dB
40AG
114,00dB
40AI
114,00dB
40AN
113,85dB
40AP
114,00dB
40AQ
114,00dB
40AR
114,00dB
40BD
114,00dB
40BE
114,00dB
40BF
114,00dB
40BP
114,00dB
Rion
UC-53
113,85dB
B&K
4144
114,00dB
4145
113,70dB
4179
113,70dB
4133
113,85dB
4134
114,00dB
4147
114,00dB
4149
113,85dB
4155
113,85dB
4165
113,85dB
4166
114,00dB
4180
113,94dB
1
4188
114,84dB
4936
114,84dB
1)
without protection grid
Содержание nor1251
Страница 2: ......
Страница 3: ...user documentation nor1251 class 1 sound calibrator ...