Norsonic nor1251 Скачать руководство пользователя страница 8

8

Using the Sound Calibrator Nor-1251

T

he Sound Calibrator Nor-1251 is a small, battery operated
sound source for the calibration of microphones and sound
measuring instruments.

The microphone is placed in an acoustic coupler where the

calibrator produces a regulated sinusoidal sound pressure sig-
nal. The calibrator is supplied with an output of 114dB SPL @
1000Hz.

Due to the principle of operation

1

, the calibration level is

virtually independent of ambient conditions like temperature,
atmospheric pressure and humidity within the specified range
of operation. The calibrator complies with IEC60942 Class1  and
has been designed to serve one-inch and smaller microphones
and sound level meters equipped with such microphones.

One-inch microphones fit directly in the calibrator coupler,

while half-inch and smaller microphones are served by applica-
tion of suitable adaptor rings. Adaptor for half-inch microphones
are included, while adaptors for other sizes must be ordered
separately.

To activate the sound calibrator:

• Press the ON button to turn on the calibrator.

The calibrator shuts off by itself after about five minutes. How-
ever, if a microphone has not been inserted or if it is removed,
the unit will switch off after a few seconds.

Your sound measuring device may or may not use a spectral

weighting network (such as the A-weighting network). For sound
calibrators with an output frequency of 1000Hz, no compensa-
tion needs to be made for this.

Operation

Using the Sound Calibrator Nor-1251 is easy and uncompli-
cated.

To calibrate with the Nor-1251, do as follows:

1. Insert the microphone to be calibrated in the coupler of the

calibrator. Use a suitable adaptor, if necessary.

2. If the background noise is high, watch the noise level and

press the ON button to switch on the calibration signal.

Verify that the calibration signal is at least 20dB above the

background noise floor.

3. Adjust the sensitivity of the microphone or measuring instru-

ment to display an excitation sound pressure level correspond-
ing to the level indicated on the calibrator itself.

When the ON button has been operated, and a microphone
inserted, the sound calibrator will stay 

on

 for approximately five

minutes. The 

on

-time will be reduced as the battery-voltage de-

creases.

Proper operation of the calibrator is indicated by the green

LED indicator – see Fig. to the right for details.

If the battery-voltage is too low to guarantee operation within

the specifications, the calibrator will switch off as soon as the

If you Have a Free-field Microphone

Most ½” free-field microphones (as defined by
IEC 61672-1) have a pressure response which gets lower
as the frequency of the input sound signal is increased.

The reason for this is that free-field microphones have
been designed to compensate for their own presence in
the sound field.

For example; when using the Norsonic microphone car-
tridge 1225, at 1000 Hz the sensitivity of your sound meas-
uring instrument should be adjusted to a value 0.15dB
lower than the nominal calibrator output level. This differ-
ence corresponds to the difference between the pres-
sure and the free-field response of the microphone – cf.
the table starting at page 15.

1)

  Patent DE 4294875 CE

Содержание nor1251

Страница 1: ... class 1 sound calibrator nor1251 A battery operated sound cali brator complying with the IEC 60942 2003 01 Class 1 producing 114 dB 1000 Hz The sound calibrator shuts off automatically when the micro phone is removed from the cou pler ...

Страница 2: ......

Страница 3: ...user documentation nor1251 class 1 sound calibrator ...

Страница 4: ...on and or the instrumentation described herein Furthermore Norsonic AS assumes no responsibility for any infringement of the intellectual property rights of third parties wherever applicable which would result from such use Norsonic AS reserves the right to amend any of the information given in this manual in order to take account of new developments If you wish to communicate with us please feel ...

Страница 5: ...eady know The manual has been divided into two parts one dealing with the use of the calibrator while the other discusses the operating principles and presents the technical specifications Depending on your requirements and your familiarisation with technical acoustics as such you may find that you use some parts of this manual often and others not at all Our main objective with this manual was to...

Страница 6: ...ed to produce a known sound pressure level when used correctly together with the measuring microphone of the sound measuring instrument All sound measuring instruments of the quality levels con sidered here feature some way of sensitivity adjustment Some have a small potentiometer that can be adjusted with a screw driver while others do it in software If the level indicated by your sound measuring...

Страница 7: ...perience that national and or international stand ards may impose restrictions on which instrument types are considered usable for a given measurement task Our problems do not end with the uncertainties of the sound measuring instruments Even sound calibrators are in fected with level uncertainties Hence these have also been divided into types called classes depending on their level ac curacy and ...

Страница 8: ...1 is easy and uncompli cated To calibrate with the Nor 1251 do as follows 1 Insert the microphone to be calibrated in the coupler of the calibrator Use a suitable adaptor if necessary 2 If the background noise is high watch the noise level and press the ON button to switch on the calibration signal Verify that the calibration signal is at least 20dB above the background noise floor 3 Adjust the se...

Страница 9: ...tself off after ap proximately five minutes Miniature loudspeaker generat ing the sound pressure level inside the coupler Silicone microphone monitoring the sound pressure level inside the coupler One side of the mi crophone diaphragm is ex posed to the sound pressure inside the coupler while the rear side of the diaphragm is located in a separate vented cavity to prevent pick up of extaneous nois...

Страница 10: ... by pulling it straight backwards 2 Pull out the battery from the battery compartment and unclip the battery connector 3 Replace the battery with a fully charged battery of the type IEC type 6LR61 nine volt alkaline 4 Alternatively a nine volt lithium battery of identical physical dimensions may be used 5 Install the battery and the rear cover See also the Fig on the previous page for details Cons...

Страница 11: ...microphone volume of 250 mm which corresponds to most measuring microphones with protecting grid mounted like the Norsonic types 1220 1225 and 1230 If the actual microphone has a volume deviating from this reference vol ume the accuracy may be improved by applying a volume correction The typical sensitivity for the level change is 0 0003 dB mm3 at 1000 Hz Hence if a microphone with an effective fr...

Страница 12: ...tem also compensates for drift in the loudspeaker The acoustic coupler is vented to the inside of the calibrator which in turn is vented to the outside for equalization to the atmospheric pressure A separate channel vents the rear side of the reference transducer to the outside of the calibrator A Light Emitting Diode LED is illuminated whenever the level control is in balance When no microphone i...

Страница 13: ...hown together with the headroom for uncertainty of measurement Nor 1251 is checked at three tempera tures 10 C 23 C and 45 C to en sure that the output level is well within the requirements set by IEC 60942 and ANSI S 1 40 over the specified tempera ture range Calibrator sound pressure level change dB as a function of temperature C 0 4dB 0dB 0 4dB 0 10 20 30 40 50 C 10 23 C 0 4dB 0dB 0 4dB Calibra...

Страница 14: ...crophone size 1 Adaptor Nor 1443 included and Adaptor Nor 1444 available separately according to IEC61094 4 Harmonic distortion Max 1 Controls Power on push button Automatic shut off CE classification EMC EN 61000 6 1 EN 61000 6 3 Safety EN 61010 1 1993 for portable equipment pollution category 2 AMBIENT REQUIREMENTS FOR SPECIFIED OPERATION Temperature range 10 to 50 C Class 1 Ambient pressure ran...

Страница 15: ...235 114 00dB 1236 114 00dB 1237 114 00dB 1245 114 00dB 1246 114 00dB Microtech MK102 113 70dB MK103 113 70dB MK202 113 85dB MK221 113 85dB MK222 113 85dB MK231 114 00dB MK250 113 85dB G R A S 40AC 113 85dB 40AD 114 00dB 40AE 113 85dB Manufacturer Microphone Type Calibration Level G R A S cont 40AF 113 85dB 40AG 114 00dB 40AI 114 00dB 40AN 113 85dB 40AP 114 00dB 40AQ 114 00dB 40AR 114 00dB 40BD 114...

Страница 16: ... 6 3 following the provisions of the LVD and EMC Directive This product has been manufactured in compliance with the provisions of the relevant internal Norsonic production stand ards All our products are tested individually before they leave the factory Calibrated equipment traceable to national and international standards has been used to carry out these tests This Declaration of Conformity does...

Отзывы: