3.
Installation requirements
Position of disconnection. The power source must be easily
reachable to plug out the device in case of emergency.
Space: a 50x50x50 cm space is required for safe and convenient installation
and operation of the Icon-5 and the Autoloader. It is recommended to place
the reagent pack behind the Icon-5 device.
For safe and unobstructed operation, leave room around the assembled
devices as indicated on the image above.
H = 500, D = 500, W = 500 (mm, minimum)
Electrical:
an additional power outlet is required for operation; the Autoloader
and the Icon-5 must be powered from the same power network.
Environment:
see specifications / operating conditions
The installation procedure requires alteration to the main device (Icon-5).
Installation must be carried out by trained professionals only
Icon-5 analyzer must run the newest software version that supports
Autoloader operation.