Norma Autoloader 

 

 

 
 
 
 
 
 
 
 
 
 
 
 

Norma Instruments Zrt. 

Papírgyár u. 58

-59, 1038 Budapest, HUNGARY 

Arany János u. 11

-13, 3530 Miskolc, HUNGARY 

 

 
 
 
 
 
 
 
 
 
 
 

PRODUCT CODE: 

NICA-AL50 

ICON-5 SW VERSION: 

1.5.1318.0 or higher 

ISSUE DATE: 

2021.03.22. 

MANUAL REVISION: 

v04

 

 

Содержание NICA-AL50

Страница 1: ...and operation Norma Autoloader Norma Instruments Zrt Pap rgy r u 58 59 1038 Budapest HUNGARY Arany J nos u 11 13 3530 Miskolc HUNGARY PRODUCT CODE NICA AL50 ICON 5 SW VERSION 1 5 1318 0 or higher ISSU...

Страница 2: ...IFICATION OF PARTS 11 FRONT VIEW 11 REAR VIEW 11 SIDE VIEW 12 TRAY 12 ASSEMBLED VIEW 13 5 INSTALLATION 14 ADJUSTMENTS 14 6 OPERATION 15 7 AUTOLOADER FUNCTIONS IN ICON 5 SOFTWARE 17 RUNNING SAMPLES USI...

Страница 3: ...The Autoloader is designed to further automate the operation of the analyzer by allowing the processing of up to 50 sample vials without operator interaction Disclaimer The manufacturer reserves the...

Страница 4: ...mpling arm Trays recognized and identified as 1 and 2 or 3 Limitations The Autoloader is inoperable on its own and can only be used with Norma Icon 5 automated hematology analyzers with support for cl...

Страница 5: ...d caps sticking higher than others this could block the movement of the pickup arm Do not move the Autoloader during operation When relocating Icon 5 equipped with Autoloader never lift the assembly b...

Страница 6: ...as potential biohazard waste always follow your local regulations for waste management Symbols on the outside of the Autoloader Symbol Meaning WARNING Risk of personal injury AUTOMATIC START UP Unexpe...

Страница 7: ...ified service personnel Your local dealer is always ready to help you in case of malfunction The contact of your local dealer is available on the insert of this Manual Please use the support normadiag...

Страница 8: ...analyzer improper use use against instructions in this manual against intended use use of non approved sampling vials damage from intentional activities overriding built in safety logic natural disast...

Страница 9: ...ge contents electrical power cord Autoloader USB 2 0 cable 4 x 2 trays for 50 vials each 4 instrument feet for Icon 5 power supply Model Sunny SYS1548 6524 T2 normaboard baseplate Grounding spare o ri...

Страница 10: ...safe and unobstructed operation leave room around the assembled devices as indicated on the image above H 500 D 500 W 500 mm minimum Electrical an additional power outlet is required for operation th...

Страница 11: ...of parts Front view Picking arm retracted position Mixer cover Tray inserted USB ports 2 slots Power button Rear view Mixer cover Picking arm retracted position Power connector Exit opening for falle...

Страница 12: ...de view Picking arm retracted position Mixing position Tray inserted Mixer cover The barcode scanner is inside Power button Tray ID pins Vial1 Vial5 Vial 6 Vial 46 Vial50 Make sure that the tray is in...

Страница 13: ...13 OM_AL_v04 Assembled view Icon 5 s closed vial opening Mixing position and barcode scanner Picking arm retracted Tray normaboard baseplate Icon 5 side Autoloader side...

Страница 14: ...sists of the following install modified instrument feet on Icon 5 install Icon 5 and Autoloader on the normaboard baseplate connect protective grounding between devices connect Autoloader power supply...

Страница 15: ...iption Status light Power off dark Power off low input voltage red with flashing blue Power off high input voltage flashing red ON all OK green In Standby no USB link to Icon 5 orange with flashing gr...

Страница 16: ...Unloading when the device finished processing the tray you can pull the tray off and load the next tray The Autoloader can identify its three trays using the ID pins on the rear of the trays Barcodes...

Страница 17: ...ons in Icon 5 software When the Icon 5 operating software detects the presence of the Autoloader an Autoloader symbol will be displayed in the upper left of the screen of Icon 5 The Autoloader symbol...

Страница 18: ...symbols Diagnostic Service mode Running No tray found Tray I loaded Tray II loaded Tray III loaded Running paused by user or due to an error Autoloader operation aborted Going to standby sleep mode A...

Страница 19: ...ence and will report the result to Icon 5 After completing the startup sequence and if both devices are in ready state detection and analysis of the samples can start Vials will be processed in rows o...

Страница 20: ...can Mode and Selected Samples Mode Selected samples mode The tray positions are displayed on the screen 1 50 Tapping on a vial position will display the sample information screen with Tray position Sa...

Страница 21: ...symbols Position selected for processing Vial back in tray waiting for results No vial was found X Vial back in tray results available Vial found Barcode OK Barcode missing Vial is in the MIXER Vial...

Страница 22: ...22 OM_AL_v04 Define individual sample information optional Tap on a vial position to access data entry screen When you are done swipe back...

Страница 23: ...canning is finished the arm will pick the vial and place it into Icon 5 s closed vial sample opening TO ICON 5 Icon 5 processes the sample AND the arm will go and test the next vial position and will...

Страница 24: ...v04 Pause Resume operation Description Status light Operation paused by User flashing yellow Tap on PAUSE or tap the power button on the Autoloader Tap on RESUME or tap the power button on the Autoloa...

Страница 25: ...h a damp lint free cloth Lift the device off from the base plate and put it in its plastic bag Place the device into the packaging box making sure it sits safely in the cushioning foam Put all accesso...

Страница 26: ...con 5 Weight 8 5 kg Autoloader only 12 5 kg including packaging Throughput 55 tests hour Power requirements External power supply universal 24 VDC 2 7 A 100 240 VAC 50 60Hz non grounded Power consumpt...

Страница 27: ...ftware to Eject the vial Arm blocked by vial during movement The vial is high or was not pushed in well enough Use supported vial types only Connection between Icon 5 and Autoloader lost Electrical co...

Страница 28: ...Vial has wrong cap or has no cap grabber cannot pick it up Use supported vial types only Vial insertion removal error grabber cannot release vial Mixer Analyzer Tray positions need to be recalibrated...

Страница 29: ...tatus icons p17 150219 CSM Vial gripper O ring replacement p25 070319 CSM v02 Actualized SW pictures All 070319 RM 090719 AGY Actualized Icon 5 model name image callout alignments All 280519 CSM 09071...