Fault finding

Fault

Possible cause

Remedy

Screen of keybord is 

not lit.

Power failure; the fuse is blown; 

the appliance is not plugged in 

correctly, the power is cut.

Check that power is connected. 

Reset the fuse

Temperature in 

freezer too high

The ventilation grille is blocked. 

The door is not closed properly. 

The quantity of items placed in 

the freezer at a time is greater 

than the capacity of the appli

-

ance. 

Thick layer of frost and ice. 

The temperature setting is too 

high

Ensure unhindered air circulation

Close the door and wait a few 

hours

Wait a few hours

Defrost the freezer.

Lower the temperature setting.

Temperature in 

freezer too low.

The temperature setting is too 

low

Raise the temperature setting

Vibration or bother-

some noise

The appliance is not level. 

Level the appliance (For model 

with adjustable feet: use a spirit 

level)

Compressor runs 

continuously.

The temperature setting is too 

low.

Too high room temperature.

Raise the temperature setting.

Ensure adequate ventilation

Display or alarm log 

shows ”RTC” error

The date and time is not set.

Please set the date and time 

Содержание ULT U100

Страница 1: ...GB Instructions for use ULT U100 ...

Страница 2: ...ear of obstruc tion WARNING Do not use other mechanical devices or other means to ac celerate the defrosting process than those recommended by the manufacturer WARNING Do not damage the refrigerant system WARNING Do not use electrical appli ances inside the refrigerated storage compartment unless they are of a type recommended by the manufacturer WARNING Do not expose the appliance to rain and sec...

Страница 3: ...n a sepa rate place and out of reach of children Before servicing or cleaning the appliance unplug the appliance from the mains or disconnect the electrical power supply If the supply cord is damaged it must be replaced by the manu facturer its service agent or similarly qualified persons in order to avoid a hazard Relevant for Australia Supply cord fitted with a plug complies with AS NZS 3112 Ple...

Страница 4: ...rial connection 7 Battery backup system at start up 8 Installation and start up 9 Maximum shelf load 10 Ventilation guards 10 Interior fitting 11 Port hole 14 Reversible door 16 Controller Operation and function 24 Defrosting cleaning and maintenance 27 Fault finding 29 Warranty sparepart and service 30 Disposal 31 ...

Страница 5: ...5 GB Lock Display Keyboard LED light Rating plate on the back side Adjustable shelf Door handle Vacuum release Adjustable feet Get to know your freezer fig 1 ...

Страница 6: ...ce The appliance is delivered in a re usable wooden box 1 Reusable wooden package 2 Cut the strapex bands or see next page if sealed with plast clips 3 Remove top 4 Remove top foam inlays 5 Remove box and lift down the appliance ...

Страница 7: ...hould be reported to the local distributor before the appliance is taken into use Clean the inside of the cabinet according to the cleaning instructions If during transport the appliance has been laid down or if it has been stored in cold surroundings colder than 5 C it must be allowed to stabilise in an upright position for at least an hour before being switched on Note If the appliance is to be ...

Страница 8: ...lled earthed socket The socket must be fused with a 10A or 13A fuse If the appliance is to be operated in a non European country check on the rating plate whether the indicated voltage and current type correspond to the values of your mains supply Data regarding voltage and obsorbed power current are given on the rating plate The power cord may be replaced by a technician only The rating plate pro...

Страница 9: ...yboard with power at power failure This makes it possible to supervise the temperatures in the unit during the power failure The battery backup system makes it possible to supervise the temperatures for 48 hours After a power failure and at the first start up the battery needs to be recharged To regain the full capacity the battery will be reloading for 10 days NOTE The battery backup system does ...

Страница 10: ...after a period of time as the underlying surface may yield under the weight of the appliance Setting up fig 4 Ventilation It is important that the appliance is well ventilated and that air can circulate freely above below and around it The figurs fig 4 6 illustrate how the neces sary air circulation around the appliance is ensured The appliance viewed from above If the appliance is placed beside a...

Страница 11: ...ance to it s final placement The guards function is to secure an absolute minimum of ventilation if the appliance is pus hed too close up against a wall Remember our guideline is minimum 100 mm free space behind the appliance to ensure suf ficient airflow 1 Locate the left Ø15 mm hole in the engi ne cover on the appliance back Push a guard in the hole 2 Then locate the identical hole in the right ...

Страница 12: ...s adjustable Demounting The shelf is demountable without use of tools Lift up to release all hooks and pull the shelf to one side Then tilt the shelf upwards in the free side and lift it out Maximum shelf load The stainless steel shelfs are designed to carry the weight of max 15 kg each 15 kg 15 kg ...

Страница 13: ...r edge Make sure the shelf direction is correct by ensuring the small punched half moon is placed in the right side 2 Place one side of the shelf hooks in the new height position 3 Gently press the hooks into the holes 4 Level the shelf and slide the hooks into the opposite holeside Press down to ensure all hooks has fallen into place ...

Страница 14: ...or compartment and remove the cover 2 Place the cover at the side of the freezer Still connected to the wire 3 The port hole is placed at the right side 4 Inside the port hole is located in the top right side and is closed with a rubber plug 5 Remove the rubber plug 6 Inside the port hole is located in right in the top 7 Remove the rubber plug ...

Страница 15: ...15 GB 8 Put in the sensor and push it all way through the tube 9 Closed the port hole by using putty 10 Close the inside hole by using putty 11 Fasten the cover with the 3 screws ...

Страница 16: ...ppliance on a table Protect top from scratches 2 Loosen and remove visible top screws use 3 mm allan key Torx 20 tool 3 Open the door to gain access 4 Remove three screws underneath the toppanel use 3 mm allan key 5 Remove lock pin use 4 mm allen key 6 Carefully pull out the toppanel ...

Страница 17: ... 2 M8 screws approx 5 mm use 5 mm allen key 9 Loosen pivot nut use 10 mm spanner 10 Lift up the top bracket pivot Pull out the door and lift it off the bottom pivot 11 Loosen door top pivot use 2 5 mm allen key if necessary 12 Remount pivot in the new hinge side ...

Страница 18: ...en key 14 Tighten pivot nut 15 Demount fix nut and top handlelock pin 16 Remount the top handlelock pin in the opposite side Tighten the fix nut again 17 Go to bottom and remove the door pivot nylon washer 18 Remove the two front feet ...

Страница 19: ...tom handlelock pin plate by loosening two screws use 4 mm allen key 20 Loosen bottom pivot nut 21 Loosen door bottom pivot 22 Remount door pivot for use in new hinge side 23 Tighten pivot 24 Remount and tighten pivot nut ...

Страница 20: ...the opposite side and rotate the bottom door bracket 27 Then remount and tighten the screws in both sides 28 Re position the door pivot nylon washer 29 Remount bottom handle lockpin plate in the opposite side by tightening the two screws again use 4 mm allen key 30 Remount and tighten the two front feet ...

Страница 21: ...orx 20 tool 32 Remove the door closer and keep it ready for mounting in the new door bottom 33 Remount the two screws without the door closer 34 Rotate the door 35 Unscrew the two doorcloser screws in this new door bottom end use Torx 20 tool 36 Place the doorcloser in the empty threadholes ...

Страница 22: ...e 39 Tighten the two M8 screws in the top bracket 40 Carefully push the toppanel back into position 41 Remount the three screws upwards in the toppanel 42 and remember the screw close to the top door pivot is mounted in hole nr 2 inwards to ensure sufficient space between the screw and the door end cap during mounting ...

Страница 23: ...23 GB 43 Remount the lock pin 44 Remount the four topscrews 45 The door re hinging operation is now complete ...

Страница 24: ...lowing table LED MODE Function ON Compressor enabled Flashing Anti short cycle delay enabled ON An alarm is occurring ON Recording activated ON Battery is fully charged Flashing Battery is beeing charged Flashing Charging problem or battery failure C F ON Measurement unit C F Flashing Programming phase ...

Страница 25: ...tton Password protected Buttons DATA How to see and modify the setpoint How to See the Set point 1 Push and immediately release the SET key the display will show the Set point value 2 Push and immediately release the SET key or wait for 60 sec to display the temperature in the unit How to Change the Set point 1 Push the SET key more than 2 sec to change the Set point value 2 The value of the set p...

Страница 26: ...rms as described in the previous chapter 1 Push the UP key once to acces the menu 2 The display shows H then push SET 3 The parameters for setting time and date occurs To set the parameter push SET push the UP or DOWN button to change the parameter Confirm by pushing SET The time and date parameters Hur hour in minutes Udy weekday dAy date on month yEA year To exit Press SET UP keys for about 10 s...

Страница 27: ...eezer Proceed as follows 1 Disconnect the power supply 2 Place a bowl of hot but not boiling water in the freezer close the door and wait ap prox 30 minutes before removing loose frost with the plastic scraper 3 To prevent water from damaging the floor during defrosting it is advisable to cover the floor around the freezer to protect against ice and splashes When all the frost and ice have melted ...

Страница 28: ...rs on the rear of the appliance be cleaned The condenser the compressor and the compressor compartment must also be kept free of dust and dirt Clean it with a vacuum cleaner regularly OBS If the appliance is not to be used for any length of time switch it off disconnect the power supply empty it clean the inside and leave the door open to allow air circula tion and prevent smells ...

Страница 29: ...e The temperature setting is too high Ensure unhindered air circulation Close the door and wait a few hours Wait a few hours Defrost the freezer Lower the temperature setting Temperature in freezer too low The temperature setting is too low Raise the temperature setting Vibration or bother some noise The appliance is not level Level the appliance For model with adjustable feet use a spirit level C...

Страница 30: ...ranty and product liability and the appliance cannot be used lawfully The approval stated on rating plate will also be invalidated Transport damage discovered by the buyer is primarily a matter to be settled between the buyer and the distributor i e the dis tributor must ensure that such complaints are resolved to the buyer s satisfaction Before calling for technical assistance please check whethe...

Страница 31: ...te waste handling For more information about collec tion and recycling of old products and batteries please contact your local municipality your waste disposal service or the point of sale where you purchased the items Penalties may be applicable for incor rect disposal of this waste in accord ance with national legislation For business users in the Euro pean Union If you wish to discard electrica...

Страница 32: ...Revision date 21 11 2016 Drawing no 80 60 234 rev 04 Reserving the right to alter specifications without prior notice Supplied by STARLAB www starlabgroup com ...

Отзывы: