Manufacturer’s Declaration

Refer to the following tables for specific information regarding this device’s compliance to IEC 60601-1-2.

Table 1: Electromagnetic Emissions

Emissions Test

Compliance

Electromagnetic Environment—Guidance

This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device 

should ensure that it is used in such an environment.

RF Emissions

CISPR 11

Group 2

This device must emit electromagnetic energy in order to perform its 
intended function. Nearby electronic equipment may be affected.

RF Emissions

CISPR 11

Class B

This device is suitable for use in all establishments, including 
domestic and those directly connected to the public low-voltage 
power supply network that supplies buildings used for domestic 
purposes.

Harmonic Emissions

IEC 61000-3-2

N/A

Voltage Fluctuations/Flicker 
Emissions

IEC 61000-3-3

N/A

Table 2: Electromagnetic Immunity

Immunity Test

IEC 60601 Test Level

Compliance 

Level

Electromagnetic Environment—Guidance

This device is intended for use in the electromagnetic environment specified below. The customer and/or user of this device should ensure that it is 

used in such an environment.

Electrostatic Discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete, or ceramic tile. If 
floors are covered with synthetic material, relative 
humidity should be at least 30%. 

Electrical Fast Transient/Burst

IEC 61000-4-4

±2 kV for power supply lines

±1 kV for input/output lines

N/A

Mains power quality should be that of a typical 
commercial or hospital environment.

Surge 

IEC 61000-4-5

±1 kV differential mode

±2 kV common mode

N/A

Mains power quality should be that of a typical 
commercial or hospital environment.

Voltage dips, short interruptions, 
and voltage variations on power 
supply input lines

IEC 61000-4-11

±5% U

T

 (>95% dip in U

T

) for 0.5 cycle

±40% U

T

 (60% dip in U

T

) for 5 cycles

±70% U

T

 (30% dip in U

T

) for 25 cycles

<5% U

T

 (>95% dip in U

T

) for 5 sec.

N/A

Mains power quality should be that of a typical 
commercial or hospital environment. 

Power Frequency (50/60 Hz) 
Magnetic Field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical 
commercial or hospital environment.

NOTE

: U

T

 is the AC mains voltage before application of the test level.

Содержание Onyx II 9560

Страница 1: ...tional hemoglobin Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following The device may not work when circulation is reduced Warm or rub th...

Страница 2: ...r assistance RF Exposure For body worn operation to maintain compliance with FCC RF exposure guidelines use only accessories that contain no metallic components and provide a separation distance of 15...

Страница 3: ...ch new Model 9560 When batteries are low the numeric displays flash once per second Remove batteries if the device will be stored for more than 30 days Replace low batteries as soon as possible using...

Страница 4: ...the startup sequence the device begins sensing the pulse indicated by the blinking pulse quality indicator Allow the device to stabilize and observe about 4 seconds of continuous green colored pulse...

Страница 5: ...as permanent damage could result 2 Dry with a soft cloth or allow to air dry Ensure that all surfaces are completely dry Testing Summary SpO2 accuracy and low perfusion testing was conducted by Nonin...

Страница 6: ...imately 600 spot checks Storage 12 months minimum Classifications per IEC 60601 1 CAN CSA C22 2 No 601 1 UL 60601 1 Degree of Protection Type BF Applied Part Enclosure Degree of Ingress Protection IP3...

Страница 7: ...l 9560 found to be within specifications Model 9560 is a precision electronic instrument and must be repaired by trained Nonin personnel only Any sign or evidence of opening the Model 9560 field servi...

Страница 8: ...netic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV cont...

Страница 9: ...20 V m 80 MHz to 2 5 GHz 20 V m where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in mete...

Отзывы: