Nonin Onyx II 9560 Скачать руководство пользователя страница 1

© 2012 Nonin Medical, Inc.
6742-001-04

MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany

Nonin and Onyx are registered trademarks of Nonin Medical, Inc.
U.S. Patents 5,490,523; 5,792,052

EC REP

Nonin Medical, Inc.

13700 1st Avenue North
Plymouth, Minnesota 55441-5443  USA

+1 (763) 553-9968 (outside US and Canada)
+46 650 401500 (Europe)
(800) 356-8874 (US and Canada)

Fax: +1 (763) 553-7807 
+46 650 401514 (Europe)

E-mail: [email protected]  
[email protected] (Europe)

www.nonin.com

Instructions for Use—English

Onyx

®

 II Model 9560 Finger Pulse Oximeter

Indications for Use

The Nonin Onyx II Model 9560 Finger Pulse Oximeter is a small, lightweight, portable, wireless device indicated for 
use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO

2

) and pulse rate of 

patients who are well or poorly perfused. It is intended for spot-checking of adult and pediatric patients on fingers 
(other than the thumb) between 0.3 – 1.0 inch (0.8 – 2.5 cm) thick. The device’s intended use environments include 
hospitals, clinics, long-term care facilities, skilled nursing facilities, emergency medical services and home 
healthcare services.

Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

Contraindications

Do not use the device in an MR environment, in an explosive atmosphere, or on infant or neonatal patients.

This device is not defibrillation proof per IEC 60601-1.

Warnings

Use the Model 9560 within its designated range (approximately 328 feet/100 meters, spherical radius, line of sight when 
connected to a class I device, from patient module to the display). Moving outside this range may cause missing, lost, and/
or inaccurate data.

This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of 
assessing clinical signs and symptoms.

The device must be able to measure the pulse properly to obtain an accurate SpO

2

 measurement. Verify that nothing is 

hindering the pulse measurement before relying on the SpO

2

 measurement. 

Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results. 

General operation of the device may be affected by the use of an electrosurgical unit (ESU).

The use of accessories other than those specified in these instructions may result in increased electromagnetic emission 
and/or decreased immunity of this device. 

This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the 
device should be observed carefully to verify normal operation.

Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.

Certain activities may pose a risk of injury, including strangulation, if the lanyard should become wrapped around your neck. 

Cautions

This device has no audible alarms and is intended only for spot-checking. 

This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may 
degrade pulse oximeter performance or affect the accuracy of the measurement include the following:

The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.

This device’s display will go blank after 30 seconds of no readings or poor readings.

In some circumstances, the device will interpret motion as good pulse quality. Minimize patient motion as much as possible.

Clean the device before applying it to a patient.

Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids into the device.

Do not use caustic or abrasive cleaning agents, or any cleaning products containing ammonium chloride or isopropyl alcohol.

A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the device’s spring. Do not 
hang the lanyard from the device’s flexible circuit.

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor.

This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems. 
This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. 
However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in health 
care and other environments, it is possible that high levels of such interference due to close proximity or strength of a source 
might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all 
equipment must be installed and put into service according to the EMC information specified in this manual. 

Portable and mobile RF communications equipment can affect medical electrical equipment.

Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 
30 days. Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at 
the same time. These actions may cause the batteries to leak.

Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device 
and device components, including batteries.

In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not 
dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor 
regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your 
distributor’s contact information.

0123

 

!

• do not apply the pulse oximeter on the 

same arm as a blood pressure cuff, 
arterial catheter or infusion line(s) (IVs)

• excessive light, such as sunlight or 

direct home lighting

• excessive motion
• moisture in the device

• improperly applied device
• finger is outside recommended size 

range

• poor pulse quality
• venous pulsations
• anemia or low hemoglobin 

concentrations

• cardiogreen and other intravascular 

dyes

• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish

Содержание Onyx II 9560

Страница 1: ...tional hemoglobin Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following The device may not work when circulation is reduced Warm or rub th...

Страница 2: ...r assistance RF Exposure For body worn operation to maintain compliance with FCC RF exposure guidelines use only accessories that contain no metallic components and provide a separation distance of 15...

Страница 3: ...ch new Model 9560 When batteries are low the numeric displays flash once per second Remove batteries if the device will be stored for more than 30 days Replace low batteries as soon as possible using...

Страница 4: ...the startup sequence the device begins sensing the pulse indicated by the blinking pulse quality indicator Allow the device to stabilize and observe about 4 seconds of continuous green colored pulse...

Страница 5: ...as permanent damage could result 2 Dry with a soft cloth or allow to air dry Ensure that all surfaces are completely dry Testing Summary SpO2 accuracy and low perfusion testing was conducted by Nonin...

Страница 6: ...imately 600 spot checks Storage 12 months minimum Classifications per IEC 60601 1 CAN CSA C22 2 No 601 1 UL 60601 1 Degree of Protection Type BF Applied Part Enclosure Degree of Ingress Protection IP3...

Страница 7: ...l 9560 found to be within specifications Model 9560 is a precision electronic instrument and must be repaired by trained Nonin personnel only Any sign or evidence of opening the Model 9560 field servi...

Страница 8: ...netic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV cont...

Страница 9: ...20 V m 80 MHz to 2 5 GHz 20 V m where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in mete...

Отзывы: