Testing Summary

SpO

2

 accuracy and low perfusion testing was conducted by NONIN

®

 Medical, Incorporated as described below.

SpO

2

 Accuracy Testing

SpO

2

 accuracy testing is conducted during induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned subjects in an independent research 

laboratory. The measured arterial hemoglobin saturation value (SpO

2

) of the sensors is compared to arterial hemoglobin oxygen (SaO

2

) value, 

determined from blood samples with a laboratory co-oximeter. The accuracy of the sensors is in comparison to the co-oximeter samples measured over 
the SpO

2

 range of 70-100%. Accuracy data is calculated using the root-mean-squared (A

rms

 value) for all subjects, per ISO 9919:2005, Standard 

Specification for Pulse Oximeters for Accuracy.

Low Perfusion Testing

This test uses an SpO

2

 Simulator to provide a simulated pulse rate, with adjustable amplitude settings of various SpO

2

 levels. The module must maintain 

accuracy in accordance with ISO 9919:2005 for pulse rate and SpO

2

 at the lowest obtainable pulse amplitude (0.3% modulation).

Specifications

Oxygen Saturation Display Range:

 

0% to 100% SpO

2

Pulse Rate Display Range:

 

18 to 321 beats per minute (bpm)

Oxygen Saturation Declared Accuracy Range: (A

rms

*): 

70% to 100% SpO

2

 ±2 digits

Low Perfusion Oxygen Saturation Declared Accuracy Range 
(A

rms

*):

70-100% SpO

2

 ±2 digits

Pulse Rate Declared Accuracy Range (A

rms

*):

20-250 BPM ±3 digits

Low Perfusion Pulse Rate Declared Accuracy Range (A

rms

*):

 

40-240 ±3 digits

Measurement Wavelengths and Output Power**: 

Red:
Infrared:

660 nanometers @ 0.8 mw maximum average

 

910 nanometers @ 1.2 mw maximum average

Temperature (Operating): 

+32° to +104°F / 0° to +40°C

Temperature (Storage/Transportation): 

-22° to +122°F / -30° to +50°C

Humidity (Operating): 

10% to 90% non-condensing 

Humidity (Storage/Transportation): 

10% to 95% non-condensing

Altitude (Operating): 

Up to 40,000 feet / 12,192 meters 

(Hyperbaric Pressure): 

Up to 4 atmospheres

Battery Life (Continuous):

Approximately 2,500 spot checks based on ~21 hours of operation using two 
AAA-size alkaline batteries, calculated at 30 seconds per spot check.

Battery Life (Storage):

4 years, minimum

Classifications per IEC 60601-1 / CSA 601.1/ UL 60601-1:

Degree of Protection: 

Type BF-Applied Part

Enclosure Degree of Ingress Protection:

IPX2

Mode of Operation: 

Continuous

This product complies with ISO 10993-1.

*± 1 A

rms

 represents approximately 68% of measurements.

 

**This information is especially useful for clinicians performing photodynamic therapy.

Warranty

NONIN

®

 MEDICAL, INCORPORATED, (NONIN

®

) warrants to the purchaser, for a period of two years from the date of purchase, each Onyx

®

 exclusive 

of the batteries, spring, carrying case, lanyard, and lanyard lock. NONIN

®

 shall repair or replace any Onyx

®

 found to be defective in accordance with this 

warranty, free of charge, for which NONIN

®

 has been notified by the purchaser by serial number that there is a defect, provided notification occurs within 

the applicable warranty period. This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any Onyx

®

 delivered to the purchaser 

which is found to be defective in any manner whether such remedies be in contract, tort or by law. This warranty excludes cost of delivery to and from 
NONIN

®

. All repaired units shall be received by the purchaser at NONIN

®

’s place of business. NONIN

®

 reserves the right to charge a fee for a warranty 

repair request on any Onyx

®

 found to be within specifications. Onyx

®

 is a precision electronic instrument and must be repaired by trained NONIN

®

 

personnel only. Any sign or evidence of opening the Onyx

®

, field service by non-NONIN

®

 personnel, tampering, or any kind of misuse of the Onyx

®

, shall 

void the warranty. All non-warranty work shall be done at NONIN

®

’s standard rates and charges in effect at the time of delivery to NONIN

®

.

Содержание Onyx II 9550

Страница 1: ...ravascular dyes may affect the accuracy of the SpO2 measurement Do not autoclave or immerse this device in liquid or use caustic or abrasive cleaning agents A flexible circuit connects the two halves Do not twist or pull the flexible circuit or overextend the device s spring Do not hang the lanyard from the device s flexible circuit strain relief A functional tester cannot be used to assess the ac...

Страница 2: ...p touches the built in stop guide 2 Make sure the finger is lying flat not on its side and is centered within the device For best results keep the Onyx II at the patient s heart or chest level 3 If the device does not turn on remove the finger and wait a few seconds before reinserting it When a finger is inserted the Onyx II performs a brief startup sequence Verify that all LEDs illuminate during ...

Страница 3: ...o 95 non condensing Altitude Operating Up to 40 000 feet 12 192 meters Hyperbaric Pressure Up to 4 atmospheres Battery Life Continuous Approximately 2 500 spot checks based on 21 hours of operation using two AAA size alkaline batteries calculated at 30 seconds per spot check Battery Life Storage 4 years minimum Classifications per IEC 60601 1 CSA 601 1 UL 60601 1 Degree of Protection Type BF Appli...

Страница 4: ... use in the electromagnetic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material relative humidity should be at least 30 Electrical Fast Transient Burst ...

Страница 5: ...is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Отзывы: